Soil pH and Salinity

SOP-L402 Revision 0.1
Author

Marc Los Huertos

EA Program, Pomona College

Published

2026-06-26

1 Purpose

This procedure describes the measurement of soil pH and electrical conductivity (EC, as a proxy for salinity) on soil-water and soil-CaCl2 extracts. Soil pH controls nutrient availability, metal solubility, microbial activity, and the fate of contaminants. Soil salinity (estimated from EC) affects plant growth, soil structure, and is a primary criterion for classifying saline, sodic, and saline-sodic soils. The methods follow Thomas (1996) for pH and Rhoades (1996) for EC, as described in the SSSA Methods of Soil Analysis, Part 3 (Rhoades, 1996; Thomas, 1996).

For meter operation, calibration, and electrode maintenance, see SOP-L300 pH and Conductivity – Bench-Top Measurements.

2 Scope

  • Applies to: Air-dried, sieved (less than 2 mm) mineral soils from all EA Program sites; compost and potting media (with appropriate extraction ratios noted in Section 6)
  • Does not apply to: In situ field pH measurements with portable probes (see SOP-F200); water sample pH and conductivity (see SOP-L300); saturated paste extraction for formal salinity classification of irrigation water
  • Location: EA Program laboratory, SGM 133

3 Definitions

  • Cation exchange capacity (CEC): The total capacity of a soil to hold exchangeable cations; closely related to soil pH buffering capacity (see SOP-L416).
  • Electrical conductivity (EC): A measure of a soil extract’s ability to conduct electricity, proportional to total dissolved ion concentration; expressed in dS m-1 or mS cm-1; used as a surrogate for soluble salt content.
  • Paste extract (saturated paste): A soil-water mixture prepared at the saturation moisture content; the USDA standard extract for measuring soil salinity.
  • pH: The negative base-10 logarithm of the hydrogen ion activity; soil pH is measured in a defined soil-to-water or soil-to-CaCl2 suspension and controls nutrient availability and microbial activity.
  • Salinity: The total concentration of dissolved salts in soil solution; high salinity reduces plant-available water and can be toxic to salt-sensitive crops; typically characterized by EC of the saturation extract (ECe).
  • Soil-to-water ratio: The ratio of soil mass to water volume used in pH and EC measurements (e.g., 1:1, 1:2, 1:5); must be reported alongside the result because pH and EC values differ between ratios.

4 Responsibilities

Role Responsibility
Lab PI Approve SOP, ensure training compliance
Lab Manager Maintain meter and electrodes, stock reagents
Trained Personnel Prepare extracts, measure and record results
New Students Complete training before independent use

5 Safety and Hazards

WarningHazards
  • Chemical: 0.01 M CaCl2 is a mild irritant. pH buffer solutions and conductivity standards are low-hazard but should not contact eyes.
  • Physical: No significant physical hazards beyond standard glassware handling.

5.1 Required PPE

5.2 Emergency Procedures

For buffer or chemical splashes, flush with water. Refer to SOP-L002 Laboratory Safety.

6 Equipment and Materials

6.1 Equipment

Item Notes
pH/conductivity meter Calibrated per SOP-L300
pH combination electrode Glass bulb with refillable KCl junction preferred
Conductivity cell With automatic temperature compensation
Analytical balance (0.01 g) See SOP-L102
Beakers (50–100 mL) Clean; one per sample plus blanks
Graduated cylinder or dispenser For volumetric delivery of water or CaCl2
Glass stirring rods One per sample or rinsed between samples
Timer For equilibration period

6.2 Reagents

Reagent Preparation / Notes
Deionized water (Type I or II) From SOP-L11
0.01 M CaCl2 solution Dissolve 1.47 g CaCl2-2H2O in 1 L DI water
pH buffer solutions (4.01, 7.00, 10.01) For pH meter calibration
Conductivity standard (1413 uS/cm) For conductivity meter calibration

7 Procedure

7.1 Sample Preparation

  1. Air-dry the soil at room temperature until constant mass (typically 24–48 hours)
  2. Pass the dried soil through a 2 mm sieve. See SOP-F400 Soil Sampling for details on sieving and homogenization.
  3. Store the sieved soil in a labeled bag or container at room temperature

7.2 Selecting the Extraction Ratio

Two standard soil-to-solution ratios are used. The choice depends on the project objectives and the need for comparability with published data:

Ratio Method When to Use
1:1 10 g soil : 10 mL water (or CaCl2) Most common in agronomic soil testing; closest to field conditions
1:5 10 g soil : 50 mL water (or CaCl2) Useful for saline soils, compost, or when more extract volume is needed for subsequent analyses
Note

Soil pH measured in 0.01 M CaCl2 is typically 0.3–0.5 units lower than pH measured in DI water because the salt displaces H+ and Al3+ from exchange sites. CaCl2 pH is less sensitive to seasonal variation in soil salt content and is preferred for monitoring studies where consistency over time matters (Thomas, 1996). Record the extraction medium (water or CaCl2) and ratio with every result.

7.3 Preparing the Soil Slurry (1:1 Method)

  1. Weigh 10.0 g of air-dried, sieved soil into a clean beaker
  2. Add 10 mL of DI water (for pHwater and EC) or 10 mL of 0.01 M CaCl2 (for pHCaCl2)
  3. Stir vigorously with a glass rod for 30 seconds
  4. Allow to stand for 10 minutes
  5. Stir briefly again before measurement

7.4 Preparing the Soil Extract (1:5 Method)

  1. Weigh 10.0 g of air-dried, sieved soil into a clean beaker
  2. Add 50 mL of DI water (or 50 mL of 0.01 M CaCl2)
  3. Stir vigorously for 30 seconds
  4. Allow to stand for 30 minutes, stirring occasionally
  5. For EC measurement on the 1:5 extract, allow the soil to settle and insert the conductivity cell into the clear supernatant. For pH, insert the electrode into the slurry while still partially suspended.

7.5 Measuring Electrical Conductivity

Important

Always measure EC before pH on the same aliquot. The pH electrode’s KCl reference junction leaks electrolyte into the sample and will artificially elevate the EC reading (APHA et al., 2012).

  1. Calibrate the conductivity meter per SOP-L300, Section 6.3
  2. Rinse the conductivity cell with DI water and blot dry
  3. Insert the cell into the supernatant (1:5 method) or into the settled portion of the slurry (1:1 method). Tap gently to dislodge air bubbles.
  4. Allow the reading to stabilize with automatic temperature compensation active
  5. Record the specific conductance (uS/cm at 25 C), sample ID, and extraction ratio
  6. Rinse the cell with DI water between samples

7.6 Measuring Soil pH

  1. Calibrate the pH meter with a three-point calibration per SOP-L300, Section 6.2
  2. Rinse the pH electrode with DI water and blot gently
  3. Insert the electrode into the slurry so that the glass bulb and reference junction are fully immersed. For the 1:1 method, the electrode should contact the settled soil.
  4. Allow the reading to stabilize (typically 30–60 seconds; may take 2–3 minutes for poorly buffered soils)
  5. Record the pH value, extraction medium (water or CaCl2), ratio, and sample ID
  6. Rinse the electrode with DI water between samples

7.7 Converting 1:5 EC to Estimated Saturated Paste EC (ECe)

The saturated paste extract (ECe) is the standard for soil salinity classification, but the 1:5 extract is quicker to prepare. An approximate conversion can be made using the following relationship, though the factor varies with texture (Rhoades, 1996):

\[ EC_e \approx EC_{1:5} \times f \]

where \(f\) is an empirical factor:

Soil Texture \(f\) (approximate)
Sand 6 – 8
Loam 8 – 10
Clay 10 – 13
Warning

These conversion factors are approximate and site-specific. For formal salinity classification (e.g., land-use planning, irrigation suitability), prepare an actual saturated paste extract per USDA Handbook 60.

8 Quality Control and Acceptance Criteria

8.1 Interpretation of Soil pH

pH Range Classification Agronomic Significance
Less than 4.5 Extremely acidic Toxic Al3+ levels; few crops tolerate
4.5 – 5.5 Strongly acidic Low P availability; Mo deficiency; lime recommended
5.5 – 6.5 Moderately acidic Suitable for most crops; some nutrient limitations
6.5 – 7.5 Neutral Optimal for most crops and nutrient availability
7.5 – 8.5 Moderately alkaline Common in arid soils; Fe, Mn, Zn may be limiting
Greater than 8.5 Strongly alkaline Possible sodicity; Na toxicity

8.2 Interpretation of Soil Salinity (ECe)

ECe (dS/m) Classification Plant Response
0 – 2 Non-saline No yield reduction for most crops
2 – 4 Slightly saline Sensitive crops affected
4 – 8 Moderately saline Many crops affected
8 – 16 Strongly saline Only tolerant crops survive
Greater than 16 Very strongly saline Few plants survive

8.3 Saline, Sodic, and Saline-Sodic Classification

Class ECe (dS/m) SAR / ESP pH
Normal Less than 4 SAR less than 13 / ESP less than 15 Variable
Saline 4 or greater SAR less than 13 / ESP less than 15 Usually less than 8.5
Sodic Less than 4 SAR 13 or greater / ESP 15 or greater Usually greater than 8.5
Saline-sodic 4 or greater SAR 13 or greater / ESP 15 or greater Variable

SAR = sodium adsorption ratio; ESP = exchangeable sodium percentage. Full SAR/ESP determination requires cation analysis beyond the scope of this SOP.

8.4 QC Requirements

  • Calibrate the pH meter before each measurement session (three-point). Acceptable slope: 90–102% of theoretical.
  • Calibrate the conductivity meter before each session
  • Re-read pH 7.00 buffer after every 10 samples; recalibrate if drift exceeds +/- 0.05 pH units
  • Re-read the conductivity standard after every 10 samples; recalibrate if drift exceeds +/- 5%
  • Run a DI water blank and a laboratory reference soil (if available) with each batch
  • Run duplicate measurements on at least 10% of samples. Acceptable precision: +/- 0.1 pH units and +/- 5% relative percent difference for EC.
  • Record extraction medium, ratio, calibration data, and QC results in the lab notebook

9 Waste Disposal

Waste Type Disposal Method
Soil slurry waste (non-hazardous) Standard trash once settled
CaCl2 extract solutions Drain with copious water (non-hazardous)
Used pH buffer solutions Collect in waste container per SOP-L300
Used conductivity standard Drain with copious water
DI water rinses Drain

10 References

  • Thomas, G.W. (1996). Soil pH and soil acidity. In: Sparks, D.L. et al. (Eds.), Methods of Soil Analysis, Part 3: Chemical Methods. SSSA Book Series No. 5. pp. 475–490.
  • Rhoades, J.D. (1996). Salinity: Electrical conductivity and total dissolved solids. In: Sparks, D.L. et al. (Eds.), Methods of Soil Analysis, Part 3: Chemical Methods. SSSA Book Series No. 5. pp. 417–435.
  • USDA. (1954). Diagnosis and Improvement of Saline and Alkali Soils. USDA Agriculture Handbook No. 60. U.S. Salinity Laboratory, Riverside, CA.
  • APHA, AWWA, WEF. (2012). Standard Methods for Examination of Water and Wastewater. 22nd ed. American Public Health Association (Eds.). Washington. 1360 pp.
APHA, AWWA, & WEF. (2012). Standard methods for the examination of water and wastewater (22nd ed., p. 1360). American Public Health Association.
Rhoades, J. D. (1996). Salinity: Electrical conductivity and total dissolved solids. In D. L. Sparks, A. L. Page, P. A. Helmke, & R. H. Loeppert (Eds.), Methods of soil analysis, part 3: Chemical methods (pp. 417–435). Soil Science Society of America.
Thomas, G. W. (1996). Soil pH and soil acidity. In D. L. Sparks, A. L. Page, P. A. Helmke, & R. H. Loeppert (Eds.), Methods of soil analysis, part 3: Chemical methods (pp. 475–490). Soil Science Society of America.

11 Training Record

Before measuring soil pH and EC independently, personnel must:

  1. Read this SOP and SOP-L300 in full
  2. Complete SOP-L002 Laboratory Safety training
  3. Demonstrate pH meter three-point calibration with acceptable slope
  4. Demonstrate conductivity meter calibration
  5. Prepare 1:1 and 1:5 extracts and measure a set of at least 5 practice soils under supervision with acceptable duplicate precision
  6. Be signed off by the PI or lab manager

12 Appendix A: Example Calculation and Reporting

Code
library(tidyverse)

# Example data entry
soil_data <- tibble(
  sample_id = c("PV-01", "PV-02", "PV-03", "PV-04", "PV-05"),
  pH_water_1_1 = c(6.82, 7.45, 5.91, 8.12, 7.03),
  pH_CaCl2_1_1 = c(6.48, 7.10, 5.55, 7.78, 6.68),
  EC_1_5_uS_cm = c(245, 1820, 85, 3450, 310),
  texture = c("loam", "clay", "sand", "clay", "loam")
)

# Estimate EC_e from 1:5 EC using texture-based conversion factors
conversion <- c(sand = 7, loam = 9, clay = 11.5)

soil_data <- soil_data %>%
  mutate(
    EC_e_dS_m = (EC_1_5_uS_cm / 1000) * conversion[texture],
    salinity_class = case_when(
      EC_e_dS_m < 2 ~ "Non-saline",
      EC_e_dS_m < 4 ~ "Slightly saline",
      EC_e_dS_m < 8 ~ "Moderately saline",
      EC_e_dS_m < 16 ~ "Strongly saline",
      TRUE ~ "Very strongly saline"
    ),
    pH_class = case_when(
      pH_water_1_1 < 4.5 ~ "Extremely acidic",
      pH_water_1_1 < 5.5 ~ "Strongly acidic",
      pH_water_1_1 < 6.5 ~ "Moderately acidic",
      pH_water_1_1 < 7.5 ~ "Neutral",
      pH_water_1_1 < 8.5 ~ "Moderately alkaline",
      TRUE ~ "Strongly alkaline"
    )
  )

print(soil_data %>% select(sample_id, pH_water_1_1, pH_CaCl2_1_1,
                           EC_e_dS_m, salinity_class, pH_class))

13 Appendix B: Reagent Preparation

13.1 M CaCl2 Solution

  1. Weigh 1.47 g of CaCl2-2H2O (MW 147.01) on the analytical balance
  2. Dissolve in approximately 800 mL of DI water in a 1 L volumetric flask
  3. Bring to 1 L with DI water and mix thoroughly
  4. Label with concentration, preparation date, and preparer’s initials
  5. Store at room temperature in a capped polyethylene or borosilicate bottle. Stable indefinitely.

14 Appendix C: Troubleshooting

Symptom Probable Cause Corrective Action
pH reading drifts continuously Poorly buffered soil; electrode junction fouled Allow longer equilibration (up to 5 min); clean electrode per SOP-L300
pH in CaCl2 higher than pH in water Measurement error; contaminated CaCl2 solution Verify CaCl2 concentration; recalibrate meter; re-measure
EC reading unexpectedly high Fertilizer residue; saline soil; contaminated beaker Verify with a blank; use acid-washed glassware (SOP-L101)
EC reading near zero Electrode not fully immersed; air bubble in cell Reposition cell; tap to dislodge bubbles
Large difference between 1:1 and 1:5 EC Normal; dilution effect Report both with the ratio clearly stated; convert to ECe using appropriate factor
pH calibration slope below 90% Aged electrode; dirty glass bulb Soak in cleaning solution or 0.1 M HCl for 15–30 min; if slope does not recover, replace electrode