Eureka Manta2 Multiprobe for Water Quality Testing
1 Purpose
This SOP describes the procedures for calibrating, operating, and maintaining the Eureka Manta2 Sub2 Multiprobe for water quality measurements including temperature, pH, depth, conductivity, and dissolved oxygen (mg/L and % saturation) (APHA et al., 2012; Eureka Environmental Engineering, 2018).
2 Scope
- Applies to: Water quality measurements in lakes, streams, rivers, and other surface water bodies
- Does not apply to: Groundwater monitoring wells (requires different deployment procedures); laboratory bench-top water analysis
- Location: All EA Program field water sampling sites
3 Responsibilities
| Role | Responsibility |
|---|---|
| Lab PI | Approve protocol, ensure training compliance |
| Field Lead | Verify calibration before deployment, coordinate |
| logistics | |
| Trained Personnel | Calibrate, operate, and clean the probe per this SOP |
| New Students | Complete training before independent use |
3.1 Required Training
Researchers using this SOP should be trained on SOP-F001 Field Safety.
4 Definitions
- Amphibian 2: The ruggedized handheld field computer (portable PC) manufactured by Eureka that controls the Manta2 probe, displays real-time readings, stores data snapshots, and connects to the probe via the underwater cable (Eureka Environmental Engineering, 2018).
- Barometric pressure: Atmospheric pressure at the measurement location, expressed in mm Hg; required as an input for dissolved oxygen calibration because oxygen solubility decreases with decreasing pressure (increasing altitude) (APHA et al., 2012).
- Calibration: The process of adjusting sensor readings to match known reference standards before field use; required before every field outing to ensure data accuracy (Eureka Environmental Engineering, 2018).
- Conductivity (specific conductance, spCond): A measure of a water sample’s ability to conduct electrical current, expressed in µS/cm or mS/cm; proportional to the concentration of dissolved ions; temperature-corrected to 25 °C for comparability across sites and seasons (APHA et al., 2012).
- Dissolved oxygen (DO): The concentration of oxygen dissolved in water, expressed as mg/L or percent saturation (%sat); measured by the HDO sensor using luminescent optical technology; decreases with increasing temperature, salinity, and altitude (APHA et al., 2012).
- HDO (Handheld Dissolved Oxygen): The luminescent optical dissolved oxygen sensor on the Manta2; measures both %sat and mg/L; calibrated in water-saturated air or with an air-saturated water standard (Eureka Environmental Engineering, 2018).
- PDOP (Position Dilution of Precision): A dimensionless measure of GPS accuracy based on satellite geometry; lower values indicate better positional accuracy; values below 3 are acceptable for most field applications.
- Reference electrode: The electrochemical half-cell that provides a stable reference potential for pH measurement; requires periodic refilling with reference electrolyte to maintain accurate pH readings (Eureka Environmental Engineering, 2018).
- Sensor Response Factor (SRF): A calibration diagnostic value indicating how the sensor responded to the reference standard; acceptable ranges are sensor-specific and indicate whether calibration was successful (Eureka Environmental Engineering, 2018).
- Snapshot: A single instantaneous record of all sensor readings captured and stored by the Amphibian at a specific date, time, and GPS location.
- Sonde (probe): The cylindrical instrument body of the Manta2 that houses all sensors (temperature, pH, conductivity, dissolved oxygen, and reference electrode) and is submerged during measurements (Eureka Environmental Engineering, 2018).
5 Safety and Hazards
- Biological/Chemical: Water bodies may be contaminated; avoid ingestion and direct skin contact.
- Physical: Slippery surfaces near water; risk of submersion; electrical components near water.
- Chemical: pH standards (pH 4, 7, 10) and conductivity standard are mild irritants.
5.1 Required PPE
Refer to SOP-F001 Field Safety for general field safety, transportation, and emergency response.
6 Equipment and Materials
| Item | Notes |
|---|---|
| Eureka Manta2 Sub2 Sonde | Inspect sensors before each use |
| Amphibian 2 field computer | Ensure fully charged (Eureka Environmental Engineering, 2018) |
| Underwater cable | Inspect for damage before use |
| pH 4, 7, and 10 standard solutions | For pH calibration; check expiration dates |
| Conductivity standard | Appropriate concentration for expected range |
| Deionized water | For rinsing sensors between calibration steps |
| Storage/calibration cup | Filled with tap water for storage |
| Field notebook | Per SOP-F003 |
7 Procedure
7.1 Startup
- Connect the underwater cable to both the sonde and the Amphibian.
- Power on the Amphibian.
- Open the Amp_2_2_6 app (or press P1 shortcut).
- Allow sensors to stabilize for 5–10 minutes before calibration.
7.2 Calibration
Calibrate before every field outing (APHA et al., 2012; Eureka Environmental Engineering, 2018). Record all calibration data (SRF values, standard concentrations, date, operator) in the field notebook.
7.2.1 pH
- Remove the storage cup. Rinse sensors with DI water.
- Submerge the sonde in pH 7 standard. Select Calibrate > pH and follow the on-screen prompts.
- Rinse with DI water. Submerge in pH 4 standard. Accept the second calibration point.
- (Optional) Verify with pH 10 standard. Record the SRF (acceptable range: 95–105).
7.2.2 Conductivity
- Rinse sensors with DI water.
- Submerge in the appropriate conductivity standard.
- Select Calibrate > Conductivity. Enter the standard value and accept (Eureka Environmental Engineering, 2018).
- Record the SRF (acceptable range: 110–120).
7.2.3 Dissolved Oxygen (HDO)
- Remove the sonde from the calibration cup. Allow the HDO sensor to equilibrate in water-saturated air for 2 minutes (sensor face moist but not submerged).
- Enter the current barometric pressure in mm Hg.
- Select Calibrate > DO and accept the 100% saturation calibration (Eureka Environmental Engineering, 2018).
- Record the SRF (should equal 100%).
7.3 Field Measurement
- Lower the sonde to the desired depth. Allow 1–2 minutes for readings to stabilize before recording.
- Press Snapshot on the Amphibian to save readings. Record the snapshot ID, location, and notes in the field notebook.
- For depth profiles, lower the sonde in 0.5 or 1 m increments, pausing 30–60 s at each depth (Eureka Environmental Engineering, 2018).
7.4 Disconnecting
- Remove the probe from the water.
- Disconnect the underwater cable from the probe.
- Insert the blue storage plug.
- Cover the cable end with the plastic protector.
8 Post-Field Procedures
8.1 Cleaning the Probe
- Clean the probe with warm soapy water (liquid dish soap or mild household cleaner).
- Clean sensor stems with a soft brush.
- Rinse thoroughly with tap water.
- Store sensors with tap water inside the cup.
8.2 Downloading Data
- Connect the Amphibian to your PC using a USB cable.
- A green window will appear with four options; select File Management.
- In the next window, navigate to the hard drive, then select My Documents.
- Locate the Excel document with your data.
- Copy the file to your project data folder.
8.3 Charging
Return the Amphibian to its charging station after each field outing.
9 Quality Control
- Calibrate all sensors before every field outing (APHA et al., 2012; Eureka Environmental Engineering, 2018).
- Verify SRF values are within acceptable ranges; do not proceed if any sensor fails calibration.
- Record a post-deployment calibration check on return to the lab; if any sensor has drifted beyond acceptable limits, flag the data.
- Compare spot readings against an independent DO meter or multiprobe where possible; readings should agree within ±0.3 mg/L for DO and ±0.1 units for pH (APHA et al., 2012).
10 Troubleshooting
| Problem | Solution |
|---|---|
| No power | Check the green LED on the Manta2; if off, the voltage is insufficient. Try another cable. |
| pH SRF out of range | Recalibrate with fresh standards. If still out of range, refill the reference electrolyte and contact the PI. |
| Conductivity SRF outside 110–120 | Recalibrate. If still out of range, contact the PI. |
| HDO %sat SRF ≠ 100 | Redo the calibration with freshly aerated water. |
| GPS not acquiring satellites | Ensure the top cap (antenna) is uncovered; move to an open area. |
11 References
13 Training Record
Before using this instrument independently, personnel must:
- Read this SOP in full
- Complete SOP-F001 Field Safety training
- Observe a trained operator perform calibration and field deployment
- Perform a supervised calibration of all sensors
- Perform a supervised field deployment with data collection
- Be signed off by the field lead or PI