Portable Spectrophotometer / Colorimeter (Field Use)
1 Purpose
This SOP describes the field operation of the Hach DR900 portable colorimeter for on-site water quality analysis (Hach Company, 2024a, 2024b). The DR900 enables rapid, field-based measurement of nutrients, metals, and other parameters using Hach pre-packaged reagent kits. Field colorimetry provides same-day results when samples cannot be transported to the lab within required hold times and supports real-time decisions during stormwater events, spill response, or reconnaissance sampling (APHA et al., 2012).
This SOP focuses on field-specific procedures (transport, outdoor operation, field QC). For detailed method chemistry, reagent handling, the Method Registry, and laboratory use of the DR3900 and DR900, see SOP-L303 Hach Colorimetric Analysis.
2 Scope
- Applies to: Field use of the Hach DR900 portable colorimeter at EA Program field sites for grab-sample analysis of nutrients (PO4, NO3, NH3), metals (Fe, Mn, Cu), and general chemistry (hardness, chloride, silica, chlorine)
- Does not apply to: Laboratory bench-top analysis (see SOP-L303); field turbidity (see SOP-F205); in situ sonde measurements (see SOP-F200)
- Location: All EA Program surface water, groundwater, and stormwater field sites
3 Responsibilities
| Role | Responsibility |
|---|---|
| PI / Field Lead | Select analytes and approve field methods |
| Trained Personnel | Operate the DR900 per this SOP; record results |
| New Students | Complete training on both this SOP and SOP-L303 before |
| independent field use |
4 Definitions
- Beer-Lambert law: The linear relationship between the absorbance of a solution and the concentration of the absorbing analyte, expressed as \(A = \varepsilon c l\), where \(\varepsilon\) is the molar absorptivity, \(c\) is concentration, and \(l\) is path length; the physical basis of colorimetric analysis (APHA et al., 2012).
- Blank (reagent blank / DI blank): A deionized water sample carried through the complete analytical procedure; used to zero the instrument and establish the baseline absorbance of the reagent system; must read at or near zero before sample measurements begin.
- Detection limit (MDL): The minimum concentration reliably distinguishable from the blank; Hach method-specific MDLs are listed in the Method Registry in SOP-L303 (Hach Company, 2024b).
- Field duplicate: A second independent measurement of the same sample, prepared through the full analytical procedure, to assess precision; RPD must be < 20% for concentrations above 5 times the detection limit.
- Reaction time: The incubation period specified in the Hach method protocol after reagent addition during which the color-forming reaction proceeds to completion; reading before reaction time has elapsed gives systematically low results (Hach Company, 2024a).
- RPD (relative percent difference): |A - B| / ((A + B) / 2) × 100; used to assess duplicate precision.
- Stored program: A factory-programmed method on the DR900 that specifies wavelength, calibration curve, and units for a specific Hach reagent kit; accessed by program number on the instrument (Hach Company, 2024a).
- Zeroing: The process of inserting a DI water blank into the DR900 sample chamber and pressing Zero to set the baseline absorbance to zero for the selected program; required before reading samples.
5 Safety and Hazards
- Chemical: Hach reagent powder pillows and liquid reagents may contain acids, oxidizers, or heavy metal indicators. Review the SDS for each kit before field use. Cadmium-containing reagents (older NitraVer 5 formulation) are hazardous waste.
- Physical: Glass sample vials can break during transport; pack in padded cases.
- Biological: Water samples may contain pathogens. Wear gloves.
5.1 Required PPE
5.2 Emergency Procedures
For reagent splashes to skin or eyes, flush with clean water for at least 15 minutes. Carry a squeeze bottle of DI water for emergency rinsing. Refer to SOP-F001 Field Safety and SOP-L003 Hazardous Materials Handling.
6 Equipment and Materials
| Item | Notes |
|---|---|
| Hach DR900 Portable Colorimeter | Charge before field day (Hach Company, 2024a) |
| Sample vials (10 mL, 25 mL, 1-inch round) | Matched to the methods being run |
| Hach reagent kits | Check expiration dates; select analytes per project |
| DI water (squeeze bottle, ≥ 500 mL) | For blanks, rinsing, and dilutions |
| Lint-free wipes (Kimwipes) | For cleaning vial exteriors |
| Silicone oil and applicator cloth | For 1-inch round vials |
| Timer | Phone timer or stopwatch for reaction times |
| Waste container (sealable) | For used reagent vials and waste solution |
| Padded carrying case | For protecting vials during transport |
| Field notebook | Per SOP-F003 |
7 Procedure
7.1 Pre-Field Preparation
- Identify the analytes to be measured; pack the corresponding reagent kits and verify none are expired.
- Pack at least 500 mL DI water and two extra clean sample vials per method.
- Charge the DR900 battery and confirm stored programs for your methods are accessible (Hach Company, 2024a).
- Print or photograph method procedure cards for each analyte (from the kit box or SOP-L303 Method Registry).
7.2 Field Setup
- Place the DR900 on a flat, stable surface out of direct sunlight. Direct sunlight on the sample chamber causes erroneous readings.
- If no shade is available, shade the instrument with a jacket or tarp during each reading.
- Turn on the DR900 and allow it to complete its self-check.
- Select the stored program for your analyte.
7.3 Blank and Calibration Check
- Fill a clean sample vial with DI water to the fill line.
- Wipe the vial exterior with a lint-free cloth.
- Insert the vial, close the lid, and press Zero.
- The display should read 0.00 (within the method detection limit); if not, clean the vial and re-zero.
- If a known mid-range standard is available, read it immediately after zeroing; the result must be within ±10% of the stated value. Record the result in the field notebook (Hach Company, 2024b).
7.4 Sample Analysis
Follow the method-specific procedure from the reagent kit or SOP-L303 Method Registry:
- Collect a grab sample per SOP-F300.
- Fill the sample vial to the fill line with sample.
- Add the reagent per method instructions.
- Cap, invert, and mix as directed. Start the timer for the required reaction time.
- When the reaction time is complete, wipe the vial exterior (apply silicone oil for round vials).
- Insert the vial into the DR900, close the lid (shield from sunlight), and press Read.
- Record the result, analyte, Hach method number, sample ID, units, date, and time in the field notebook.
- Rinse the vial three times with DI water before the next sample.
7.5 Waste Management
- Collect all used reagent packets, vials, and waste solution in a sealable container.
- Do not discard reagent waste in the field.
- Return waste to the lab and dispose of per SOP-L003.
8 Quality Control
- Zero the instrument with DI water before each batch of samples.
- Run one DI blank and one mid-range standard at the start of each field session; record results in the field notebook (Hach Company, 2024b).
- Analyze one field duplicate per 10 samples (minimum one per event); RPD must be < 20% for concentrations above 5 × the detection limit.
- If the field blank reads above the MDL, investigate contamination (dirty vials, contaminated DI water) and re-run.
- Note ambient temperature; some Hach methods are temperature-sensitive. If the sample is outside the method-specified range (typically 20–25 °C), note the potential bias (Hach Company, 2024a).
9 References
11 Training Record
Before using the DR900 in the field independently, personnel must:
- Read this SOP and SOP-L303 in full
- Demonstrate a complete field analysis (zero, standard check, sample analysis, waste collection) under supervision
- Be signed off by the PI or field lead