Purpose
This SOP establishes the overarching quality assurance (QA) and quality control (QC) framework for all analytical work in the EA Program laboratory. It defines the minimum QC requirements that apply across methods, describes how to calculate and interpret QC metrics, specifies detection limit procedures, and establishes the data documentation and reproducibility practices required for every analysis. Individual analytical SOPs reference this document for general QC principles and add method-specific requirements (APHA et al., 2012; U.S. Environmental Protection Agency, 2007).
Quality assurance is the system of management actions that ensures the analytical process produces results of known and documented quality. Quality control is the set of technical activities (blanks, standards, duplicates, spikes) embedded in each analytical batch to measure and control that quality in real time (Taylor, 1987).
Scope
- Applies to: All quantitative chemical, physical, and biological analyses performed in the EA Program laboratory, including those described in SOPs L002 through L999 and analyses of field-collected samples (SOPs F001 through F999)
- Does not apply to: Qualitative screening tests (e.g., visual soil texture by feel, SOP-F401) unless those tests include a quantitative component
- Location: EA Program laboratory, SGM 133
Responsibilities
| Lab PI (QA Manager) |
Approve this SOP; review QC data quarterly; authorize corrective actions |
| Lab Manager |
Maintain instrument logs, reagent inventory, and QC records; train new analysts |
| Analyst |
Implement all QC requirements for each batch; flag QC failures; document corrective actions |
| Data Reviewer |
Verify QC compliance before data are reported or published |
Definitions
- Accuracy: The closeness of a measured value to the true or accepted reference value; assessed using certified reference materials (CRMs) and laboratory-fortified blanks (LFBs) (Taylor, 1987).
- Calibration blank: A solution containing only the analytical solvent or reagent matrix (no analyte); used to establish the instrument baseline (zero) for each calibration sequence.
- Certified reference material (CRM): A material with certified analyte concentrations traceable to a national or international metrology institute (NIST, USGS, IRMM); used to verify analytical accuracy across instruments and over time (APHA et al., 2012).
- Control chart (Shewhart chart): A time-series plot of QC sample results with control limits (typically ±2σ warning limits and ±3σ action limits); used to distinguish random analytical variation from systematic drift or instrument malfunction.
- Continuing calibration verification (CVS): A mid-range calibration standard analyzed at regular intervals (typically every 10 samples) during an analytical run; must recover within ±10% of the stated value; failure requires re-calibration and re-analysis of affected samples (APHA et al., 2012).
- Detection limit – Method Detection Limit (MDL): The minimum concentration of an analyte that can be detected with 99% confidence under stated method conditions; determined from at least 7 replicate analyses of a low-level spike following 40 CFR Part 136 Appendix B (U.S. Environmental Protection Agency, 2016).
- Detection limit – Limit of Blank (LOB): The highest signal expected from a blank sample; LOB = meanblank + 1.645 × SDblank (Armbruster & Pry, 2008).
- Detection limit – Limit of Detection (LOD): The lowest analyte concentration that can be reliably distinguished from the LOB; LOD = LOB + 1.645 × SDlow-concentration; represents 95% detection probability (Armbruster & Pry, 2008).
- Detection limit – Limit of Quantitation (LOQ): The lowest concentration at which the analyte can be measured with acceptable precision and accuracy, typically defined as 10 × SD at the LOD spike concentration; also known as the practical quantitation limit (PQL) (Armbruster & Pry, 2008).
- Field blank: A sample of reagent-grade water processed through the entire field collection and preservation procedure; detects contamination introduced during sampling and transport.
- Laboratory-fortified blank (LFB): A reagent water blank spiked with a known concentration of target analytes and processed through the full analytical procedure; monitors ongoing accuracy and extraction efficiency.
- Laboratory-fortified matrix (LFM): A field sample spiked with a known concentration of target analytes; detects matrix-specific suppression or enhancement of the analytical signal.
- Method blank: A reagent water sample carried through the full analytical procedure alongside field samples; detects contamination introduced by reagents, equipment, or laboratory environment (APHA et al., 2012).
- Precision: The reproducibility of repeated measurements of the same sample; expressed as relative percent difference (RPD) for duplicates or relative standard deviation (RSD) for replicate analyses (Taylor, 1987).
- Relative percent difference (RPD): |A - B| / ((A + B) / 2) × 100; the standard measure of duplicate precision; typical acceptance criteria are RPD ≤ 10–20% depending on method and matrix.
- Spike recovery (%): (Measured spiked concentration – Unspiked concentration) / Added concentration × 100; typical acceptance range is 80–120% for most environmental matrices (APHA et al., 2012).
Safety and Hazards
QC samples contain the same analytes as field samples and are subject to the same hazard considerations as the parent analytical method. Consult the relevant analytical SOP for safety information.
General QC Requirements
Every Analytical Batch
Every analytical batch must include at minimum (APHA et al., 2012; U.S. Environmental Protection Agency, 2007):
| Method blank |
1 per batch |
Detect laboratory contamination |
| Calibration blank |
Start of each calibration |
Instrument baseline |
| CVS (check standard) |
Every 10 samples |
Detect calibration drift |
| Duplicate |
10% of samples |
Assess precision |
| LFB |
1 per batch |
Assess accuracy |
Additional Elements (as applicable)
| Field blank |
Each sampling event |
| Matrix spike (LFM) |
5% of samples or 1 per batch |
| CRM |
1 per batch (where certified material is available) |
| Equipment blank |
When non-dedicated sampling equipment is used |
Acceptance Criteria
| Method blank |
< MDL (analyte absent) |
Investigate contamination; do not report affected samples |
| CVS |
±10% of stated value |
Re-calibrate; re-analyze affected samples |
| Duplicate RPD |
≤ 10–20% (method-specific) |
Flag results; investigate and re-analyze if warranted |
| LFB recovery |
80–120% |
Flag results; check reagents and instrument |
| LFM recovery |
70–130% |
Suspect matrix interference; consult PI |
| CRM recovery |
Within certified uncertainty interval |
Recalibrate; re-analyze |
Acceptance criteria in individual analytical SOPs take precedence over these general values when they are stricter or when regulatory requirements apply.
Detection Limit Determination
Method Detection Limit (MDL)
The MDL is determined following 40 CFR Part 136 Appendix B (U.S. Environmental Protection Agency, 2016):
- Select a spike concentration at 1–5× the estimated MDL.
- Prepare and analyze at least 7 replicates of the spiked solution through the full analytical method.
- Calculate the standard deviation (\(s\)) of the replicates.
- Multiply by the one-tailed \(t\)-value at 99% confidence (\(t_{n-1, 0.99}\); for \(n=7\): \(t = 3.143\)):
\[\text{MDL} = t_{(n-1,\, 0.99)} \times s\]
- If the spike concentration is outside the range of 0.5–20× the computed MDL, repeat with an adjusted spike concentration (U.S. Environmental Protection Agency, 2016).
Limit of Detection and Limit of Quantitation
The LOB, LOD, and LOQ hierarchy follows Armbruster & Pry (2008):
\[\text{LOB} = \mu_{\text{blank}} + 1.645\,\sigma_{\text{blank}}\]
\[\text{LOD} = \text{LOB} + 1.645\,\sigma_{\text{low}}\]
\[\text{LOQ} = 10\,\sigma_{\text{low}}\]
where \(\sigma_{\text{low}}\) is the standard deviation of replicate analyses at the low-level spike concentration. For right-censored environmental data, apply the non-parametric methods described in Helsel (2012).
Control Charts
Maintain control charts for each analyte/method combination to track long-term analytical performance:
- Plot CRM, LFB, or CVS recovery for each batch on a time-series chart.
- Calculate and draw lines for:
- Center line: Mean recovery over the last 20+ batches
- Warning limits: ±2 standard deviations from the mean
- Action limits: ±3 standard deviations from the mean
- A result between warning and action limits is a warning; investigate but do not necessarily re-analyze.
- A result beyond the action limits requires corrective action and re-analysis.
- Systematic trends (7+ consecutive results on one side of the center line, or a steady drift) require investigation even if individual points are within limits (Taylor, 1987).
Corrective Actions
When a QC failure occurs:
- Stop the analytical run; do not report affected samples.
- Identify the probable cause (contaminated blank, degraded standard, instrument malfunction, matrix interference).
- Correct the problem (replace reagents, recalibrate, perform maintenance).
- Re-run the affected samples with fresh QC elements.
- Verify that the repeated run passes all QC criteria.
- Document the failure, root cause, corrective action, and re-analysis results in the QC log and lab notebook.
QC Requirements by Method Category
| Calibration blank |
Each cal |
Each cal |
Each cal |
N/A |
Each cal |
| Method blank |
1/batch |
1/batch |
1/batch |
1/batch |
1/batch |
| Field blank |
1/event |
N/A |
N/A |
N/A |
N/A |
| Duplicate |
10% |
10% |
10% |
10% |
10% |
| LFB |
1/batch |
1/batch |
1/batch |
N/A |
1/batch |
| Matrix spike |
5% (or 1/batch) |
5% |
5% |
N/A |
5% |
| CRM |
1/batch (if avail.) |
1/batch |
1/batch |
Positive control |
1/batch |
| CVS |
Every 10 samples |
Every 10 |
Every 10 |
N/A |
Every 10 |
Waste Disposal
QC samples generate the same waste streams as routine samples. Dispose of QC waste (blanks, standards, spikes) according to the parent method’s waste disposal procedure.
References
- APHA et al. (2012)
- Taylor (1987)
- U.S. Environmental Protection Agency (2007)
- U.S. Environmental Protection Agency (2016)
- Armbruster & Pry (2008)
- Helsel (2012)
- Gandrud (2015)
- Xie (2015)
APHA, AWWA, & WEF. (2012). Standard methods for the examination of water and wastewater (22nd ed., p. 1360). American Public Health Association.
Armbruster, D. A., & Pry, T. (2008). Limit of blank, limit of detection and limit of quantitation. Clinical Biochemistry Reviews, 29(Suppl 1), S49–S52.
Gandrud, C. (2015). Reproducible research with R and RStudio (2nd ed.). CRC Press.
Helsel, D. R. (2012).
Statistics for censored environmental data using Minitab and R (2nd ed.). John Wiley & Sons.
https://doi.org/10.1002/9781118162729
Taylor, J. K. (1987). Quality assurance of chemical measurements. Lewis Publishers.
U.S. Environmental Protection Agency. (2007). Guidance for quality assurance project plans (EPA/240/R-02/009). Office of Environmental Information.
U.S. Environmental Protection Agency. (2016).
Definition and procedure for the determination of the method detection limit, revision 2 (40 CFR Part 136, Appendix B). U.S. Environmental Protection Agency.
https://www.epa.gov/cwa-methods/mdl-determination-procedure
Xie, Y. (2015). Dynamic documents with R and knitr (2nd ed.). CRC Press.
Training Record
Before applying QC frameworks independently, personnel must:
- Read this SOP in full
- Demonstrate correct MDL calculation using the calc_mdl() function on a test dataset
- Prepare a complete QC summary table for one analytical batch under supervision
- Set up and interpret a control chart for one analyte/method combination
- Be signed off by the PI or lab manager