Preparing a Wet Mount for Microscopy
1 Purpose
This procedure describes how to prepare a wet mount slide for light microscopy. A wet mount suspends a specimen in a thin layer of water (or other mounting medium) beneath a cover slip, preserving the natural shape and motility of living cells and small organisms. It is the standard preparation method for examining fresh biological samples under bright field or phase contrast illumination (Murphy & Davidson, 2012).
2 Scope
- Applies to: Preparation of temporary wet mount slides for examination of algae, protozoa, bacteria, fungi, plant cells, soil microorganisms, and other small specimens in the EA Program laboratory
- Does not apply to: Permanent slide preparation (fixed, stained, and sealed mounts); dry mounts; hanging-drop preparations for motility studies; fluorescence specimen preparation (see SOP-L211)
- Location: EA Program laboratory, SGM 133
3 Responsibilities
| Role | Responsibility |
|---|---|
| Lab PI | Approve SOP, ensure training compliance |
| Trained Personnel | Prepare slides following this procedure |
| New Students | Complete training before independent use |
4 Definitions
- Cover slip (coverslip): A thin (0.13–0.17 mm) square or circular piece of optical glass placed over the specimen on a microscope slide; standardizes the optical path length and protects the objective from liquid samples; the EA Program standard is No. 1.5 (0.17 mm), matched to the objectives on the Echo Revolve microscope (Murphy & Davidson, 2012).
- Immersion medium: Liquid placed between the specimen and the coverslip to improve optical contact, reduce refraction artifacts, and support live organisms during observation; options include water, glycerol, or immersion oil depending on the application.
- Live mount: A wet mount prepared without fixation or staining; allows observation of living organisms and motile behavior; must be examined quickly before evaporation or oxygen depletion affects the specimen.
- Mounting medium: The liquid in which the specimen is suspended; distilled or deionized water is standard; saline, glycerol, or specific staining solutions may be used depending on the application (Murphy & Davidson, 2012).
- Refraction artifact: Optical distortion caused by differences in refractive index between the sample, mounting medium, coverslip, and objective; minimized by matching refractive indices and maintaining thin, uniform wet mounts.
- Wet mount: A temporary microscopy preparation method in which a specimen is suspended in a drop of liquid (usually water) on a glass slide and covered with a coverslip; the standard preparation for aquatic microorganisms, algae, macroinvertebrates, and other specimens examined under transmitted light (Murphy & Davidson, 2012).
5 Safety and Hazards
- Physical: Microscope slides and cover slips are fragile glass. Broken glass can cause cuts. Handle by the edges and dispose of chipped or cracked glass in the broken-glass box.
- Biological: Environmental samples (water, soil, compost) may contain unknown microorganisms. Treat all samples as potentially infectious and wash hands after handling.
5.1 Required PPE
5.2 Emergency Procedures
For cuts from broken glass, apply direct pressure and rinse with water. Seek first aid if bleeding is significant. Refer to SOP-L002 Laboratory Safety.
6 Required Materials
| Item | Catalog No. | Notes |
|---|---|---|
| Microscope slides (75 × 25 mm) | TBD | Standard glass, pre-cleaned |
| Cover slips (No. 1.5, 22 × 22 mm) | TBD | 0.17 mm thickness; matched to objectives on the Revolve microscope (Murphy & Davidson, 2012) |
| Squeeze bottle with DI water | Or distilled water | |
| Disposable transfer pipettes or eyedroppers | For placing water drops | |
| Forceps (fine-tipped) | For handling specimens and cover slips | |
| Toothpicks | For lowering cover slips | |
| Lint-free wipes (Kimwipes) | For blotting excess water |
The original SOP flagged that catalog numbers for slides and cover slips should be added. Update this table when part numbers are confirmed with the lab manager.
7 Estimated Time
This procedure requires approximately 5 minutes per slide.
8 Procedure
8.1 Placing the Specimen
- Clean a microscope slide with a lint-free wipe to remove dust and fingerprints.
- Using forceps, place the specimen in the center of the slide. For liquid samples (pond water, culture broth), skip to Step 3 and pipette the sample directly.
8.2 Adding the Mounting Liquid
- Using a transfer pipette or eyedropper, place one to two drops of water onto the specimen. Adjust the volume based on specimen size: larger or thicker specimens may need slightly more water, while very small specimens need less.
8.3 Applying the Cover Slip
- Pick up a cover slip by its edges with forceps or clean, dry fingers. Never touch the flat surfaces.
- Place one edge of the cover slip on the slide adjacent to the specimen, holding it at approximately a 45-degree angle.
- Slowly lower the cover slip using the tip of a toothpick or forceps, allowing the water to spread gradually beneath the glass. Lowering at an angle lets air escape and minimizes bubble formation (Murphy & Davidson, 2012).
Air bubbles distort the image and can be mistaken for cellular structures. If bubbles are present after placing the cover slip, gently tap the cover slip with the eraser end of a pencil or press lightly with a toothpick to work bubbles toward the edge. If large bubbles persist, remove the cover slip, add a fresh drop of water, and try again.
8.4 Adjusting Water Volume
- Too much water: If the cover slip floats or slides freely, the water layer is too thick. Hold the edge of a lint-free wipe against the edge of the cover slip and let capillary action draw off excess water until the cover slip sits flat.
- Too little water: If dry areas are visible beneath the cover slip, place a small drop of water at the edge and let capillary action draw it underneath. Do not lift the cover slip.
8.5 Examining the Slide
- Place the prepared slide on the microscope stage and examine following SOP-L211 Echo Labs Revolve Microscope. Start at low magnification (4× or 10×) and increase as needed.
- For unstained, transparent specimens, phase contrast mode typically provides better visibility than bright field (see SOP-L211, Section 6.5).
9 Quality Control
- Verify that the cover slip lies flat and is not tilted by air bubbles or excess sample material.
- Confirm that the water layer is thin enough for the specimen to remain in a single focal plane at high magnification.
- If examining motile organisms, note that excessive pressure from the cover slip can restrict movement; use slightly more water to provide adequate space.
10 Waste Disposal
| Waste Type | Disposal Method |
|---|---|
| Used glass slides and cover slips | Broken-glass box |
| Used toothpicks and wipes | Standard trash |
| Biological specimen waste | Biohazard bag if potentially infectious; otherwise standard trash |
11 Troubleshooting
| Problem | Probable Cause | Corrective Action |
|---|---|---|
| Air bubbles under cover slip | Cover slip lowered too quickly | Remove cover slip, re-wet, lower at 45° slowly |
| Cover slip slides around | Too much water | Blot excess with lint-free wipe |
| Dry patches under cover slip | Too little water | Add a drop at the edge; capillary action fills the gap |
| Specimen drifts out of view | Water current under cover slip | Wait for flow to stop; blot if excess water is present |
| Image blurry at high magnification | Thick water layer or dirty slide | Blot excess water; clean slide and re-prepare |
12 References
- Murphy & Davidson (2012)
14 Training Record
Before preparing wet mount slides independently, personnel must:
- Read this SOP in full
- Complete SOP-L002 Laboratory Safety training
- Complete SOP-L211 Revolve Microscope training
- Prepare at least two wet mount slides under supervision and demonstrate proper cover slip technique (no air bubbles, correct water volume)
- Be signed off by the PI or lab manager