Soil Organic Matter – Loss on Ignition
1 Purpose
This SOP describes the estimation of soil organic matter (SOM) content using the loss-on-ignition (LOI) method. SOM is quantified by measuring the mass lost when oven-dried soil is combusted at 550 °C in a muffle furnace (Heiri et al., 2001; Nelson & Sommers, 1996).
2 Scope
- Applies to: Air-dried, sieved soil samples collected per SOP-F400 Soil Sampling and Processing.
- Does not apply to: Soils with high calcium carbonate content (CaCO3 decomposes above ~450 °C and may inflate LOI values; consider an alternative method such as elemental analysis for calcareous soils (Heiri et al., 2001)).
- Location: EA Program laboratory, SGM 133.
3 Definitions
- Ash-free dry mass (AFDM): The mass remaining after removing water (drying at 105 °C) and organic matter (igniting at 550 °C); equivalent to the inorganic mineral mass of the sample.
- Crucible: A heat-resistant porcelain or ceramic container used to hold soil samples during furnace ignition; must be pre-ignited and weighed to constant mass before use.
- Inorganic carbon (IC): Carbon associated with carbonates (CaCO3, MgCO3); released as CO2 during ignition above ~550 °C; can inflate LOI values in calcareous soils if ignition temperature is not controlled or if acid pre-treatment is omitted (Heiri et al., 2001).
- Loss on ignition (LOI): The percentage mass loss when an oven-dried soil is ignited at 550 °C for 4 hours: \(LOI = (M_{105} - M_{550}) / M_{105} \times 100\); used as a proxy for SOM content (Heiri et al., 2001).
- Soil organic matter (SOM): The organic fraction of soil, including plant and animal residues at various stages of decomposition, microbial biomass, and stabilized humus; estimated by LOI as approximately \(\text{SOM} \approx 1.72 \times \text{LOI}\) (the Van Bemmelen factor) when organic carbon is the target (Nelson & Sommers, 1996).
- Tare mass: The pre-weighed mass of the empty, pre-ignited crucible; subtracted from all subsequent measurements to obtain sample mass.
- Van Bemmelen factor: The empirical conversion factor (1.724) relating organic carbon content to SOM by assuming that SOM contains 58% carbon by mass; use with caution for soils with unusual organic matter composition (Nelson & Sommers, 1996).
4 Responsibilities
| Role | Responsibility |
|---|---|
| Lab PI | Approve protocol, ensure training compliance |
| Trained Personnel | Perform analysis, record and report results |
| New Students | Complete training before independent use |
5 Safety and Hazards
- Thermal: Muffle furnace operates at 550 °C; severe burn risk. Drying oven operates at 105 °C. Ceramic crucibles remain extremely hot for many minutes after removal from the furnace.
- Physical: Crucible tongs can slip; always use two hands when moving hot crucibles.
5.1 Required PPE
5.2 Emergency Procedures
Refer to SOP-L002 Laboratory Safety and SOP-L111 Muffle Furnace. For burns, cool with running water for at least 10 minutes and seek medical attention.
6 Equipment and Materials
6.1 Equipment
| Item | Notes |
|---|---|
| Muffle furnace | Capable of 550 ±10 °C; programmable ramp preferred |
| Drying oven | Regulated at 105 ±2 °C |
| Analytical balance | Resolution ≥ 0.0001 g (0.1 mg) |
| Desiccator | With active color-indicating silica gel |
| Crucible tongs | Heat-resistant; long-handled |
6.2 Consumables
| Item | Notes |
|---|---|
| Porcelain or ceramic crucibles | Pre-ignited to constant mass |
| Weighing paper or spatulas | For soil transfer |
7 Procedure
7.1 Crucible Preparation
- Pre-ignite crucibles in the muffle furnace at 550 °C for at least 1 hour to remove any residual organic matter.
- Transfer to a desiccator and cool to room temperature.
- Weigh each crucible to the nearest 0.0001 g and record as the tare mass (\(W_{\text{crucible}}\)).
7.2 Oven Drying at 105 °C
- Using a spatula, transfer 5–10 g of air-dried, sieved soil into a pre-weighed crucible.
- Record the mass of crucible plus moist or air-dry soil.
- Place uncovered crucibles in the drying oven at 105 °C for at least 24 hours.
- Transfer crucibles to a desiccator and cool to room temperature (approximately 20–30 minutes).
- Weigh to the nearest 0.0001 g and record as \(W_{\text{105}}\) (mass of oven-dried soil only, i.e., crucible + dry soil mass minus crucible tare).
7.3 Combustion at 550 °C
- Transfer the crucibles containing oven-dried soil directly into the muffle furnace (Heiri et al., 2001).
- Ramp the furnace to 550 °C (if programmable, use a ramp rate of approximately 10 °C per minute to prevent crucible cracking and reduce convective loss of fine particles).
- Hold at 550 °C for 4 hours (Heiri et al., 2001).
- Turn off the furnace and allow crucibles to cool inside the furnace until they can be safely transferred (below approximately 200 °C).
- Transfer crucibles to a desiccator and cool to room temperature.
- Weigh to the nearest 0.0001 g and record as \(W_{\text{550}}\) (mass of combusted soil, excluding crucible).
7.4 Calculation
Soil organic matter content is calculated as (Heiri et al., 2001):
\[ \text{SOM (\%)} = \left( \frac{W_{105} - W_{550}}{W_{105}} \right) \times 100 \]
where:
- \(W_{105}\) = mass of oven-dried soil (g)
- \(W_{550}\) = mass of soil after combustion at 550 °C (g)
To convert LOI (%) to approximate organic carbon (%OC), divide by 1.724 (the Van Bemmelen factor); note that this conversion assumes constant elemental composition and is less accurate for samples with unusual organic matter (Nelson & Sommers, 1996).
8 Quality Control
- Run analytical duplicates for at least 10% of samples; coefficient of variation between duplicates should be less than 5% (Heiri et al., 2001).
- Include an empty crucible blank in each furnace run to verify no mass gain from ambient contamination.
- Include a laboratory reference soil with known SOM content in each batch.
- Verify oven (105 °C) and furnace (550 °C) temperatures using a calibrated thermocouple at regular intervals.
9 Waste Disposal
| Waste Type | Disposal Method |
|---|---|
| Combusted soil residue | Standard trash (inert mineral ash) |
| Crucibles | Clean and reuse |
10 References
- Heiri et al. (2001)
- Nelson & Sommers (1996)
- SOP-L002 Laboratory Safety
- SOP-L111 Muffle Furnace
12 Training Record
Before performing this procedure independently, personnel must:
- Read this SOP in full
- Complete SOP-L111 Muffle Furnace training
- Demonstrate safe operation of the muffle furnace and drying oven
- Demonstrate proper use of the analytical balance and desiccator
- Perform the procedure under supervision with acceptable duplicate precision
- Be signed off by the PI or lab manager