This procedure describes the determination of alkalinity in water samples by potentiometric titration with standardized sulfuric acid, following Standard Methods 2320 B (APHA et al., 2012). Alkalinity is the acid-neutralizing capacity of water and is primarily a function of carbonate, bicarbonate, and hydroxide content. It is a fundamental parameter for characterizing water chemistry, assessing buffering capacity, and interpreting corrosivity, hardness, and suitability for irrigation or aquatic life.
2 Scope
Applies to: Surface water, groundwater, stormwater, and wastewater samples collected per SOP-F300
Does not apply to: Soil alkalinity, industrial process water requiring specialized titration protocols, or samples with alkalinity above 500 mg/L as CaCO3 (which require 0.1 N titrant; see note in Section 6.4)
Location: EA Program laboratory, SGM 133
3 Responsibilities
Role
Responsibility
Lab PI
Approve SOP, ensure training compliance
Lab Manager
Prepare and standardize titrant, maintain pH meter
Trained Personnel
Perform titrations, record and report results
New Students
Complete training before independent use
4 Definitions
Acid-neutralizing capacity (ANC): The capacity of a water to neutralize strong acid; for most natural waters, ANC is approximately equal to total alkalinity.
Alkalinity: The capacity of water to resist acidification, expressed in mg/L as CaCO3; in natural waters, primarily contributed by bicarbonate (HCO3-), carbonate (CO32-), and hydroxide (OH-).
Bicarbonate alkalinity: Alkalinity attributable to HCO3-; the dominant form in most fresh waters at pH 6.3–10.3.
Endpoint pH: The pH at which the titration is stopped to correspond to the equivalence point of the reaction between the alkalinity species and the standard acid; 4.5 for routine total alkalinity, 8.3 for phenolphthalein (P) alkalinity.
Normality (N): The number of gram-equivalents of a substance per liter of solution; for H2SO4 used in alkalinity titrations, 1 N = 0.5 M because each H2SO4 donates 2 protons.
Phenolphthalein alkalinity (P alkalinity): Alkalinity measured to the pH 8.3 endpoint; represents OH- and half of CO32- content; zero if the sample pH is initially below 8.3.
Potentiometric titration: Titration in which the endpoint is determined by monitoring electrode potential (pH) rather than by a color indicator.
Standardization: The process of determining the exact concentration of a titrant by titrating a primary standard of known purity and mass (e.g., anhydrous Na2CO3).
Total alkalinity (T alkalinity): Alkalinity measured to the pH 4.3–4.5 endpoint; represents the sum of all alkalinity species.
5 Safety and Hazards
WarningHazards
Chemical: Sulfuric acid (0.02 N) is a dilute acid but can cause skin and eye irritation. Concentrated H2SO4 used to prepare stock solutions is severely corrosive.
Physical: Glassware (burets, beakers) poses a breakage and laceration risk.
5.1 Required PPE
5.2 Emergency Procedures
For acid splashes to skin or eyes, flush immediately with water at the eyewash station for at least 15 minutes. For concentrated acid spills, neutralize with sodium bicarbonate and refer to SOP-L003 Hazardous Materials Handling.
Prepare from concentrated H2SO4 or purchase pre-made; standardize against Na2CO3 (see Section 6.1)
Standard H2SO4, 0.02 N
Dilute from 0.1 N stock (see Section 6.2)
Sodium carbonate (Na2CO3), primary standard grade
For standardizing the acid titrant
pH buffer solutions (pH 4.01, 7.00, 10.01)
For pH meter calibration
Deionized water (Type I or II)
For reagent preparation, blanks, and rinsing
6.3 Optional Indicator Solutions
Color indicators may be used for preliminary or routine titrations when the sample is free of interfering color and turbidity. Potentiometric (pH meter) titration is preferred for accuracy.
Indicator
Endpoint pH
Application
Phenolphthalein
8.3
Phenolphthalein (P) alkalinity
Bromcresol green
4.5
Total (T) alkalinity
Mixed bromcresol green – methyl red
4.5
Total (T) alkalinity
7 Procedure
7.1 Standardizing the 0.1 N H2SO4
Dry 3–5 g of primary standard Na2CO3 at 250 C for 4 hours and cool in a desiccator
Weigh 2.500 +/- 0.200 g (to the nearest 0.001 g) into a 1 L volumetric flask. Dissolve in DI water and bring to volume. This produces an approximately 0.05 N Na2CO3 solution.
Pipette 40.00 mL of the Na2CO3 solution into a 250 mL beaker. Add approximately 60 mL of DI water.
Titrate potentiometrically with the approximately 0.1 N H2SO4 to a pH of about 5
Remove the electrodes, rinse into the beaker, and gently boil the solution for 3–5 minutes under a watch glass to expel CO2
Cool to room temperature, rinse the watch glass into the beaker, and continue titrating to the pH inflection point
where \(A\) = g Na2CO3 weighed into the 1 L flask, \(B\) = mL Na2CO3 solution taken for titration, and \(C\) = mL acid used.
Adjust to 0.1000 N or use the measured normality in all subsequent calculations. At 0.1000 N, 1.00 mL = 5.00 mg CaCO3.
7.2 Preparing the 0.02 N H2SO4 Working Titrant
Dilute 200.00 mL of standardized 0.1000 N H2SO4 to 1000 mL with DI water in a volumetric flask
Mix thoroughly
Verify the normality by titrating 15.00 mL of 0.05 N Na2CO3 as described in Section 6.1
At 0.02 N, 1.00 mL = 1.00 mg CaCO3
Store in a tightly capped borosilicate glass or polyethylene bottle. Do not keep for longer than 1 week without re-standardization.
7.3 Sample Collection and Handling
Collect samples in polyethylene or borosilicate glass bottles per SOP-F300
Fill sample bottles completely with no headspace; CO2 exchange with air will alter the alkalinity
Analyze samples as soon as possible after collection, ideally within 24 hours. Refrigerate at 4 C if analysis must be delayed.
Do not filter, dilute, concentrate, or otherwise alter the sample before titration
7.4 Selecting Sample Size and Titrant
Choose the sample volume and titrant normality to use a reasonable volume of titrant (ideally 20 mL or more from a 50 mL buret) for good volumetric precision:
Expected Alkalinity (mg/L as CaCO3)
Sample Volume
Titrant Normality
Less than 20
200 mL
0.02 N (10 mL microburet)
20 – 100
100 mL
0.02 N
100 – 500
100 mL
0.1 N
If the expected alkalinity is unknown, perform a preliminary titration with indicator to estimate the range before running the potentiometric titration.
7.5 pH Meter Calibration
Calibrate the pH meter with a three-point calibration (pH 4.01, 7.00, 10.01) per SOP-L300 before each titration session. Verify that the calibration slope is between 90% and 102% of theoretical.
7.6 Potentiometric Titration
Measure the selected sample volume (e.g., 100 mL) into a clean beaker using a graduated cylinder
Place the beaker on the magnetic stir plate and add a clean stir bar. Set the stirrer to a speed that produces a gentle vortex without splashing or entraining air.
Immerse the pH electrode in the sample. Record the initial pH.
Fill the buret with standardized H2SO4 (0.02 N or 0.1 N as selected in Section 6.4). Record the initial buret reading.
Phenolphthalein alkalinity (P): If the initial pH is above 8.3, titrate to pH 8.3, recording the volume of acid used. This is the P alkalinity endpoint.
Total alkalinity (T): Continue titrating to the total alkalinity endpoint. Select the endpoint pH from the table below based on the expected alkalinity concentration:
Total Alkalinity (mg/L as CaCO3)
Endpoint pH
Approximately 30
4.9
Approximately 150
4.6
Approximately 500
4.3
Routine or automated analyses
4.5
As the endpoint is approached, add acid in smaller increments (0.1–0.2 mL) and allow pH equilibrium to be reached before adding more titrant
Record the final buret reading
TipTip
For best results, record the pH after each addition of acid and construct a full titration curve. The inflection point (steepest slope) on the curve confirms the endpoint pH. This is especially important for samples with unusual chemistry (e.g., high phosphate or silicate content) where the standard endpoint pH may not correspond to the true equivalence point.
7.7 Low-Alkalinity Method (Below 20 mg/L as CaCO3)
For samples with alkalinity below 20 mg/L, use a 200 mL sample volume and titrate with 0.02 N H2SO4 from a 10 mL microburet. Record intermediate pH values to construct a titration curve and determine the endpoint from the inflection point. Report only low-alkalinity values determined by this method.
8 Quality Control and Acceptance Criteria
8.1 Calculations
Total alkalinity (mg/L as CaCO3):
\[\text{Alkalinity} = \frac{A \times N \times 50{,}000}{V}\]
where \(A\) = mL of standard acid used, \(N\) = normality of the acid, and \(V\) = mL of sample.
Phenolphthalein alkalinity uses the same formula but substitutes the volume of acid titrated to pH 8.3.
8.2 Alkalinity Species
Once P and T alkalinity are determined, the individual carbonate species can be calculated:
Condition
Hydroxide (OH-)
Carbonate (CO32-)
Bicarbonate (HCO3-)
P = 0
0
0
T
P < T/2
0
2P
T – 2P
P = T/2
0
2P
0
P > T/2
2P – T
2(T – P)
0
P = T
T
0
0
All values are expressed as mg/L CaCO3.
8.3 QC Requirements
Standardize titrant before each batch and at least weekly
Run a DI water blank at the start of each session; the blank should consume less than 0.2 mL of 0.02 N acid
Prepare and titrate a laboratory fortified blank (LFB) using a known concentration of Na2CO3 at least once per batch of 20 samples. Recovery should be 90–110%.
Run duplicate titrations on at least 10% of samples. Acceptable precision: +/- 5% relative percent difference for samples above 20 mg/L as CaCO3.
Record all QC results in the laboratory notebook
9 Waste Disposal
Waste Type
Disposal Method
Titrated samples (no hazardous additions)
Drain with copious water
Spent acid titrant
Neutralize with sodium bicarbonate to pH 6–8, then drain
APHA, AWWA, WEF. (2012). Standard Methods 2320 B: Alkalinity – Titration Method. In: Standard Methods for Examination of Water and Wastewater. 22nd ed. American Public Health Association (Eds.). Washington. 1360 pp.
Add approximately 900 mL of DI water to a 1 L volumetric flask
Slowly add 2.8 mL of concentrated H2SO4 to the water (always add acid to water)
Mix, cool to room temperature, and bring to volume with DI water
Standardize against primary standard Na2CO3 (Section 6.1)
12.2 N H2SO4 from 0.1 N Stock
Transfer 200.00 mL of standardized 0.1000 N H2SO4 to a 1 L volumetric flask
Bring to volume with DI water and mix thoroughly
13 Appendix B: Example Calculation
Code
# Alkalinity calculation per Standard Methods 2320 Bacid_mL <-24.5# mL of standard acid usedacid_N <-0.02# Normality of acid (eq/L)sample_mL <-100# mL of sample titratedalkalinity_mg_L <- (acid_mL * acid_N *50000) / sample_mLcat(sprintf("Total alkalinity: %.1f mg/L as CaCO3\n", alkalinity_mg_L))
14 Appendix C: Troubleshooting
Symptom
Probable Cause
Corrective Action
Endpoint difficult to identify
Low alkalinity; flat titration curve
Use 200 mL sample, 0.02 N acid, microburet; construct full titration curve
Blank consumes excessive acid
CO2 absorbed into DI water; contaminated glassware
Use freshly boiled and cooled DI water; acid-wash glassware
Duplicate precision exceeds 5% RPD
Inconsistent titrant delivery; pH meter drift
Check buret for air bubbles; recalibrate pH meter; ensure pH equilibrium at endpoint
Alkalinity result seems too high
Sample exposed to air (CO2 loss raised pH)
Analyze samples promptly; fill bottles with no headspace
Alkalinity result seems too low
CO2 absorbed during storage or titration
Minimize exposure to air; do not stir vigorously enough to entrain air