NanoDrop One Spectrophotometer
1 Purpose
This procedure describes the operation, calibration verification, and maintenance of the Thermo Scientific NanoDrop One micro-volume UV-Vis spectrophotometer. The NanoDrop measures nucleic acid and protein concentrations from samples as small as 1–2 uL and provides absorbance ratios (A260/A280 and A260/A230) used to assess sample purity. For the role of NanoDrop measurements within the DNA extraction workflow, see SOP-L700 DNA Extraction.
2 Scope
- Applies to: All use of the NanoDrop One spectrophotometer in the EA Program laboratory, including nucleic acid quantification, protein quantification (A280 method), and custom wavelength measurements
- Does not apply to: Cuvette-based spectrophotometers (see SOP-L201), fluorometric quantification (see SOP-L213), or Hach benchtop spectrophotometry (see SOP-L303)
- Location: Instrument bench, Seaver ??
3 Responsibilities
| Role | Responsibility |
|---|---|
| Lab PI | Approve SOP, authorize instrument access |
| Lab Manager | Schedule maintenance, manage software updates |
| Trained Personnel | Follow this SOP, report malfunctions |
| New Students | Complete training before independent use |
4 Definitions
- A260: Absorbance at 260 nm; the wavelength at which double-stranded DNA and RNA absorb most strongly; used to quantify nucleic acid concentration.
- A280: Absorbance at 280 nm; absorbed primarily by aromatic amino acids (tyrosine, tryptophan) and by phenolic contaminants; used to assess protein contamination in nucleic acid preparations.
- A260/A280 ratio: The ratio of absorbance at 260 nm to 280 nm; pure dsDNA gives ~1.8; pure RNA gives ~2.0; values below 1.7 indicate protein or phenol contamination.
- A260/A230 ratio: The ratio of absorbance at 260 nm to 230 nm; values of 2.0–2.2 are expected for pure nucleic acids; low values indicate contamination with chaotropic salts (guanidinium), carbohydrates, or humic acids.
- Micro-volume spectrophotometry: Absorbance measurement using 1–2 µL of sample held by surface tension between a pedestal and an optical fiber arm; eliminates the need for cuvettes and allows measurement of very small volumes.
- Pedestal: The fixed lower optic of the NanoDrop on which the sample droplet is placed; must be cleaned with a lint-free wipe between measurements to prevent carryover.
- Pathlength normalization: Automatic adjustment of the measured absorbance to a standardized 10 mm pathlength equivalent; allows direct comparison with traditional cuvette measurements.
5 Safety and Hazards
- Optical: The NanoDrop uses a pulsed xenon flash lamp. Do not look directly into the light source when the pedestal arm is open.
- Biological: Samples may contain hazardous biological material. Treat all samples as potentially infectious.
- Chemical: Cleaning solvents (e.g., 0.5 M HCl for decontamination) require appropriate handling.
5.1 Required PPE
5.2 Emergency Procedures
For biological spills on or near the instrument, clean and disinfect per SOP-L002 Laboratory Safety. If liquid enters the instrument housing, disconnect power immediately and contact the lab manager.
6 Equipment and Materials
6.1 Equipment
| Item | Location | Notes |
|---|---|---|
| NanoDrop One spectrophotometer | Instrument bench, SGM 133 | Thermo Scientific |
| Connected laptop or NanoDrop display | Instrument bench | For data export |
6.2 Consumables
| Item | Notes |
|---|---|
| Lint-free wipes (Kimtech) | For pedestal cleaning between samples |
| Nuclease-free water | For blanking and pedestal cleaning |
| Elution buffer | Must match the buffer used in extraction |
7 Procedure
7.1 Startup
- Open the NanoDrop One software (or wake the integrated touchscreen)
- Select the appropriate application (e.g., Nucleic Acid, Protein A280, or UV-Vis)
- Clean the upper and lower pedestals with 2 uL of nuclease-free water; wipe with a lint-free tissue
- Perform a blank measurement (see Section 6.2)
7.2 Blanking
- Pipette 2 uL of the blank solution onto the lower pedestal. The blank must match the sample matrix – use the same elution buffer or solvent that the samples are dissolved in.
- Lower the sampling arm and tap Blank
- Verify that the blank spectrum is flat and the baseline absorbance at 260 nm is near zero
- Wipe both pedestals with a lint-free tissue
Re-blank whenever you switch sample types, change buffers, or if the instrument has been idle for more than 30 minutes.
7.3 Measuring Samples
- Pipette 2 uL of sample onto the lower pedestal
- Lower the sampling arm and tap Measure
- Record the following for each sample:
- Concentration (ng/uL)
- A260/A280 ratio
- A260/A230 ratio
- Wipe both pedestals with a lint-free tissue between every sample
- Repeat for all samples
7.4 Interpreting Results
| Parameter | Expected Range | Interpretation |
|---|---|---|
| A260/A280 | 1.8–2.0 | Values below 1.8 suggest protein or phenol contamination |
| A260/A230 | 2.0–2.2 | Values below 2.0 suggest humic acid, carbohydrate, salt, or guanidine contamination |
For environmental DNA extracts, A260/A230 ratios are frequently below 2.0. If the NanoDrop concentration is more than twice the Qubit reading, the UV absorbance is likely inflated by co-extracted contaminants; use the Qubit value for planning downstream reactions.
7.5 Data Export
- Save the session file (.tsv or .csv) to the designated project folder on the connected laptop
- Use a consistent file-naming convention:
YYYY-MM-DD_NanoDrop_ProjectName.tsv - Record results in the lab notebook and/or project data sheet
7.6 Shutdown and Daily Maintenance
- Wipe both pedestals with 2 uL of nuclease-free water, then dry with a lint-free tissue
- If protein or concentrated DNA residue is visible, clean with 2 uL of 0.5 M HCl, wait 10 seconds, then rinse with nuclease-free water
- Close the software and lower the sampling arm
- Cover the instrument with the dust cover if available
8 Quality Control and Acceptance Criteria
- Verify instrument performance at the start of each session by measuring a blank; the 260 nm absorbance should read 0.00 +/- 0.02
- If the NanoDrop and Qubit concentrations disagree by more than a factor of two, the NanoDrop reading is likely inflated by contaminants; trust the Qubit value
- TBD – establish a schedule for periodic performance verification using Thermo Scientific NanoDrop verification standards
9 Waste Disposal
| Waste Type | Disposal Method |
|---|---|
| Used lint-free wipes | Standard trash |
| Pipette tips | Standard trash (or biohazard if samples are hazardous) |
| Residual sample on pedestal | Wiped and cleaned as part of shutdown |
10 References
- Thermo Fisher Scientific. NanoDrop One/OneC User Guide. Thermo Fisher Scientific, Waltham, MA.
- Desjardins, P. & Bhatt, T. (2010). Micro-volume quantification of nucleic acids. Current Protocols, A.3J.1–A.3J.16.
11 Training Record
Before using the NanoDrop independently, personnel must:
- Read this SOP in full
- Complete SOP-L002 Laboratory Safety training
- Observe a trained operator demonstrate blanking, measurement, and data export
- Perform a practice session (blank plus at least 5 samples) under supervision
- Be signed off by the PI or lab manager