Purpose
This procedure describes the assembly, operation, and maintenance of vacuum filtration systems used in the EA Program laboratory. Vacuum filtration separates suspended solids from liquids by drawing the liquid through a filter membrane or glass fiber filter under reduced pressure. It is used across many analytical methods, including soil and water extract filtration (SOPs L403, L409, L418), microplastics analysis (SOP-L308), chlorophyll extraction (SOP-L310), and eDNA concentration (SOP-L700).
Scope
- Applies to: All vacuum filtration in EA Program laboratories using filter flasks, filter funnels (Buchner, glass frit, analytical, or Nalgene disposable units), and Fisher Scientific Maxidry vacuum pumps or building vacuum lines
- Does not apply to: Syringe filtration (push-through, no vacuum); centrifugal ultrafiltration; field filtration of eDNA samples (covered in SOP-F700)
- Location: SGM 133 and SGM 134
Definitions
- Filter funnel: The upper component of a vacuum filtration assembly that holds the filter membrane and directs liquid onto it; typically borosilicate glass or polysulfone.
- Filtrate: The liquid that passes through the filter membrane and collects in the side-arm flask; the fraction of interest in most dissolved-analyte analyses.
- Glass fiber filter (GFF): A depth filter composed of borosilicate glass microfibers; commonly 0.7 µm nominal pore size (GF/F); used for chlorophyll, particulate carbon, and TSS analyses.
- Membrane filter: A thin polymeric film with a defined pore size (typically 0.45 µm); used to separate dissolved from particulate fractions in water samples.
- Nominal pore size: The manufacturer-stated pore size of a filter; actual retention varies by manufacturer and depends on the particle size distribution of the material being filtered.
- Retentate: The material retained on the filter membrane; the fraction of interest for TSS, chlorophyll, and particulate analyses.
- Side-arm (vacuum) flask: A thick-walled borosilicate flask with a side arm for connection to the vacuum line; designed to withstand the pressure differential during filtration.
- Vacuum trap: A secondary flask or bottle interposed between the filtration funnel and the vacuum pump to catch filtrate overflow and protect the pump.
Responsibilities
| Lab PI |
Approve SOP |
| Lab Manager |
Maintain pumps, replace tubing |
| Trained Personnel |
Operate filtration systems per this SOP |
| New Students |
Complete training before independent use |
Safety and Hazards
- Glassware under vacuum: Filter flasks (Erlenmeyer with sidearm) are under reduced pressure during operation. Cracked or starred flasks can implode. Inspect all glassware before each use; reject any with visible damage.
- Chemical: Filtrates may contain hazardous chemicals (acids, solvents). Handle filtrate according to the analytical SOP.
- Biological: Samples may contain pathogens. Wear gloves.
- Sharp edges: Broken glass fiber filters and fritted glass funnels can cut skin. Handle with forceps.
Emergency Procedures
If a filter flask implodes, do not touch the broken glass. Sweep up fragments with a brush and dustpan (not hands) and dispose in a sharps container or labeled “Broken Glass” box. If contaminated with hazardous chemicals, follow SOP-L004 Chemical Spill Response.
Equipment and Materials
Filtration Hardware
| Fisher Scientific Maxidry vacuum pump |
4 |
SGM 134 |
Oil-free diaphragm pump |
| Filter flask (sidearm Erlenmeyer, 500 mL or 1 L) |
Multiple |
SGM 134 |
Borosilicate glass; inspect for cracks |
| Rubber stopper with hole (matched to funnel stem) |
Multiple |
SGM 134 |
Connects funnel to flask |
| Vacuum tubing (thick-wall rubber or silicone) |
– |
SGM 134 |
Check for cracks and kinks |
| In-line vacuum trap (flask or bottle between the filter flask and pump) |
1 per pump |
SGM 134 |
Prevents liquid from reaching the pump |
Funnel Types
| Buchner funnel (porcelain or polypropylene) |
General filtration with flat filter paper or GF filters |
| Glass frit funnel (sintered glass, various porosities) |
Fine filtration; reusable; requires careful cleaning |
| Nalgene analytical filter unit (disposable, sterile, 0.45 um PES) |
Sterile filtration for microbiological and eDNA work |
| Millipore glass filter holder (47 mm) |
Membrane filtration for water quality (TSS, chlorophyll, microplastics) |
Procedure
Assembly
- Inspect the filter flask for cracks or stars. Do not use damaged flasks.
- Connect vacuum tubing from the flask sidearm to the vacuum trap inlet, and from the vacuum trap outlet to the pump
- Place the rubber stopper into the neck of the flask
- Seat the appropriate funnel into the stopper
- If using a 47 mm filter holder: assemble the base, place the filter on the fritted support using forceps (avoid touching the filter with fingers), and clamp the funnel top in place
Pre-Wetting or Pre-Combustion (If Required)
- For gravimetric analyses (TSS): pre-dry and pre-weigh the filter per the analytical SOP
- For DOC filtration: pre-rinse the filter with 50–100 mL of sample or DI water and discard the first rinse (removes filter-leached carbon)
- For pre-combusted GF/F filters (chlorophyll, DOC): filters should have been fired at 450 C for 4 hours per SOP-L111 and stored in a clean container
Filtration
- Turn on the vacuum pump. Verify suction by briefly covering the funnel opening with a gloved hand (you should feel resistance).
- Pour the sample into the funnel slowly and steadily. For particulate-laden samples, add in small aliquots to avoid overloading the filter.
- Allow the liquid to pass through completely
- Rinse the funnel walls with a small volume of DI water (or the matrix specified in the analytical SOP) to transfer any adhering particles to the filter
- Continue vacuum for 10–30 seconds after the last visible liquid has passed through to remove residual moisture (do not over-dry if the filter will be extracted for chlorophyll or DOC)
- Turn off the vacuum pump
- Carefully disconnect the funnel from the stopper
Filter Handling
- Remove the filter from the funnel using clean forceps
- Transfer the filter to the appropriate container per the analytical SOP:
- Aluminum tin (for drying/weighing: TSS)
- Amber vial with extraction solvent (chlorophyll)
- Extraction tube (eDNA, microplastics)
- Petri dish (if storing before analysis)
- Label the container immediately
Filtrate Handling
- If the filtrate is the analyte of interest (e.g., dissolved nutrients, DOC), transfer it from the filter flask to a labeled sample vial using a clean graduated cylinder or by pouring
- Acidify or preserve the filtrate per the analytical SOP
- If the filtrate is waste, dispose per SOP-L006
Cleanup
- Discard used disposable funnels and filters per the analytical SOP
- Rinse reusable funnels, stoppers, and filter flasks with DI water immediately after use
- For glass frit funnels: back-flush with DI water, then soak in an appropriate cleaning solution (acid bath for inorganic work, detergent for general use) per SOP-L101
- Coil vacuum tubing loosely (do not kink) and store
Vacuum Pump Maintenance
| Check vacuum tubing for cracks/kinks |
Before each use |
| Empty and clean the vacuum trap |
After each use session |
| Verify pump suction (should reach at least 20 in Hg / 500 mmHg) |
Monthly |
| Replace vacuum tubing |
Annually or when degraded |
| Service pump (diaphragm replacement) |
Per manufacturer schedule or when suction declines |
Always use a vacuum trap (secondary flask or bottle) between the filter flask and the pump. If liquid reaches the pump, it can damage the diaphragm and contaminate the pump internals.
Quality Control
- For analytical methods requiring pre-weighed filters (TSS), follow the weighing QC in the analytical SOP
- For DOC filtration, verify filter blanks by filtering DI water and analyzing the filtrate; blank DOC should be less than 0.5 mg/L
- Inspect all glassware for cracks before every use
Waste Disposal
| Used filters (non-hazardous) |
Standard trash |
| Used filters (hazardous matrix) |
Chemical waste per SOP-L006 |
| Filtrate (hazardous) |
Appropriate waste stream per SOP-L006 |
| Filtrate (non-hazardous) |
Drain with water if approved by EH&S |
| Broken glassware |
Sharps container or “Broken Glass” box |
References
- Fisher Scientific. Maxidry Vacuum Pump User Manual.
- APHA, AWWA, WEF. (2012). Standard Methods for Examination of Water and Wastewater. 22nd ed. Section 2540: Solids.
Training Record
Before operating vacuum filtration systems independently, personnel must:
- Read this SOP in full
- Demonstrate proper assembly of the filtration system (flask, stopper, funnel, trap, pump, tubing)
- Demonstrate correct filter handling with forceps
- Perform one filtration under supervision
- Be signed off by the PI or lab manager