Using a Vortex Mixer
1 Purpose
This procedure describes the operation and care of vortex mixers used in the EA Program laboratory. Vortex mixers produce a rapid orbital or oscillating motion that creates a vortex inside a tube or small container, providing fast and thorough mixing of liquids, suspensions, and reagents. They are used routinely in DNA extraction (SOP-L700), Qubit sample preparation (SOP-L213), chemical extractions, and general reagent mixing.
2 Scope
- Applies to: All vortex mixers in the EA Program laboratory, including the Fisher Scientific vortex mixers (SGM 134, qty 3) and the Vortex-Genie 2 (SGM 133)
- Does not apply to: Orbital shakers (for extended, gentle agitation of multiple samples); magnetic stir plates; sonicators (see SOP-L115)
- Location: SGM 133 and SGM 134
3 Responsibilities
| Role | Responsibility |
|---|---|
| Lab PI | Approve SOP |
| Lab Manager | Maintain equipment, replace cup attachments |
| Trained Personnel | Operate the vortex mixer per this SOP |
| New Students | Complete training before independent use |
4 Definitions
- Continuous mode: A vortex mixer operating mode in which the platform spins continuously when the power switch is on; used when mixing many tubes in sequence without holding each one.
- Touch mode: A vortex mixer operating mode in which the platform spins only when downward pressure is applied to the platform; allows the user to control mixing duration for each tube individually.
- Vortex: A spiral or whirling motion of liquid produced when a tube’s closed end is pressed off-center against a spinning platform; efficient for resuspending pellets, dissolving solids, and homogenizing small volumes.
- Vortex mixer (Vortex-Genie): A benchtop mixer with an eccentric rotating cup or platform that creates a vortex in tubes pressed against it; operated at variable speeds (typically 0–3,200 rpm).
5 Safety and Hazards
- Aerosol generation: Vortexing uncapped or loosely capped tubes can spray liquid and generate aerosols. Always cap tubes securely before vortexing. If the sample contains hazardous biological or chemical material, vortex inside a fume hood or biological safety cabinet.
- Tube ejection: At high speed, improperly held tubes can slip from the cup head and be thrown from the mixer. Hold tubes firmly against the cup or use the hands-free cup attachment.
- Breakage: Thin-walled glass tubes may shatter under vigorous vortexing. Use polypropylene or thick-walled borosilicate tubes whenever possible.
- Entanglement: Keep hands, hair, jewelry, and loose clothing away from the spinning cup head.
- Electrical: Do not operate the mixer with a damaged power cord or loose components. Inspect the cord before use; report damage to the lab manager.
5.1 Required PPE
6 Equipment
| Item | Location | Notes |
|---|---|---|
| Fisher Scientific vortex mixer | SGM 134 | Qty 3; touch or continuous mode |
| Vortex-Genie 2 (Scientific Industries) | SGM 133 | Variable speed; touch or continuous mode |
| Cup head (standard) | Attached | For single tubes (1.5 mL to 50 mL) |
| Flat head or platform head (optional) | SGM 134 | For microplates or multiple small tubes held in a rack |
7 Procedure
7.1 Pre-Use Inspection
- Verify the vortex mixer is clean and free from visible damage
- Ensure the power cord is intact and properly connected
- Check that the rubber cup head or attachment is securely installed and not cracked or deformed
7.2 Setup
- Verify that the vortex mixer is on a stable, flat bench surface and that the rubber feet are intact (prevents the unit from walking during operation)
- Confirm that the appropriate head is installed:
- Cup head: for individual tubes (microcentrifuge tubes, 15 mL and 50 mL centrifuge tubes, glass test tubes)
- Platform head: for microplates or tube racks (if available)
- Set the speed dial to the desired setting. For most routine mixing, a medium speed (setting 5–7 on the Vortex-Genie 2 dial) is sufficient. Use lower speeds for gentle mixing and higher speeds for vigorous vortexing (e.g., bead disruption steps in DNA extraction).
7.3 Operation – Touch Mode
Touch mode is the default for most applications. The mixer runs only while the tube is pressed against the cup head.
- Cap the tube securely
- Hold the tube at a slight angle (approximately 15–30 degrees from vertical) and press the bottom of the tube firmly into the center of the cup head
- The motor engages automatically. A visible vortex should form in the liquid.
- Vortex for the duration specified in the analytical SOP (typically 2–5 seconds for brief mixing, 10–30 seconds for thorough resuspension)
- Release the tube from the cup to stop
7.4 Operation – Continuous Mode
Use continuous mode for hands-free operation or when extended vortexing is required.
- Switch the mode selector to Continuous (or On)
- Adjust the speed dial to the desired setting
- Place the capped tube in the cup head or in a tube holder secured to the platform head
- The motor runs continuously at the set speed
- After the prescribed duration, switch back to Touch mode or turn off the unit
- Remove the tube
7.5 Vortexing Multiple Tubes
For rapid sequential vortexing of many tubes (e.g., a rack of Qubit assay tubes):
- Set the mixer to touch mode at the desired speed
- Work through the rack systematically, pressing each tube against the cup for 2–3 seconds
- Return each tube to the rack before picking up the next
For simultaneous vortexing of a plate or rack, attach the platform head and secure the plate or rack with the provided clamp or tape. Run in continuous mode for the prescribed duration.
7.6 Special Applications
| Application | Speed Setting | Duration | Notes |
|---|---|---|---|
| Brief reagent mixing | Medium (5–7) | 2–5 s | Touch mode; just to mix |
| Qubit assay tube mixing (L213) | Medium (5–7) | 2–3 s | Brief vortex after adding sample to working solution |
| DNA extraction bead disruption (L700) | Maximum (10) | Per kit protocol (often 10 min on adapter) | Use continuous mode with horizontal adapter if available |
| Resuspending pellets after centrifugation | High (8–10) | 10–30 s | Until pellet is fully resuspended |
| Mixing soil extracts (L403, L409) | Medium (5–7) | 5–10 s | Ensure homogeneous suspension before filtering |
8 Maintenance
| Task | Frequency |
|---|---|
| Wipe the cup head and body with 70% ethanol | After each use session |
| For eDNA/PCR work: wipe with 10% bleach, then ethanol | After each use session |
| Check rubber feet for wear | Monthly |
| Inspect cup head for cracks or deformation | Monthly |
| Replace cup head | When cracked or deformed |
If liquid spills into the cup head or body of the mixer, unplug the unit immediately, wipe up the spill, and allow to dry completely before re-use. Liquid inside the motor housing can cause electrical failure.
9 Waste Disposal
No waste is generated by the vortex mixer itself. Dispose of sample waste per the relevant analytical SOP and SOP-L006 Waste Disposal.
10 References
- Scientific Industries. Vortex-Genie 2 Instruction Manual. Scientific Industries, Bohemia, NY.
12 Appendix A: Troubleshooting
| Symptom | Probable Cause | Corrective Action |
|---|---|---|
| Mixer does not start | Power disconnected or cord damaged | Check power supply; inspect cord; try another outlet |
| Excessive vibration or walking | Unit on uneven surface; rubber feet worn | Reposition on a flat surface; replace rubber feet |
| Poor or incomplete mixing | Speed too low; tube not centered on cup | Increase speed; press tube firmly into center of cup |
| Sample leakage during vortexing | Tube cap not secured; tube overfilled | Cap tightly; reduce fill volume to leave headspace |
| Abnormal noise or grinding | Cup head loose or damaged; internal bearing wear | Tighten or replace cup head; report to lab manager for service |
| Liquid entered motor housing | Spill from uncapped tube | Unplug immediately; wipe and dry thoroughly; do not use until fully dry |