Pygmy and Type AA Velocity Meters
1 Purpose
This SOP describes the specifications, assembly, maintenance, and ordering of replacement parts for the Price Pygmy and Price Type AA mechanical current meters used by the EA Program. These vertical-axis cup-type meters convert water velocity into rotational counts and are the standard instruments for measuring point velocities during stream discharge measurements (Rantz et al., 1982) (see also SOP-F301 Stream Discharge Measurement).
2 Scope
- Applies to: Price Pygmy current meters and Price Type AA current meters owned by the EA Program, including standard cat-whisker and magnetic-head contact chamber variants
- Does not apply to: Electronic velocity meters (e.g., Marsh-McBirney, FlowTracker ADV), propeller-type meters (e.g., Ott), or acoustic Doppler instruments
- Location: Instruments stored in the EA field equipment room; maintenance performed in SGM 133
3 Responsibilities
| Role | Responsibility |
|---|---|
| Lab PI | Approve purchases, authorize meter retirement |
| Field Lead | Perform pre-field spin tests, coordinate maintenance |
| Trained Personnel | Assemble, inspect, and clean meters per this SOP |
| New Students | Complete training before handling meters |
4 Definitions
- AquaCalc: A digital flow computer that counts current meter revolutions electronically and converts them to velocity automatically using the stored rating table; eliminates manual timing and reduces transcription error.
- Cat-whisker contact: A spring-wire contact element in the Type AA meter’s contact chamber; makes electrical contact with a solder bead on the shaft once per revolution (single-revolution contact) or once per five revolutions (penta contact); must be checked for wear and proper contact force before each field season (Rantz et al., 1982).
- Contact chamber: The sealed housing on the Type AA meter that contains the electrical contact mechanism (cat-whisker wire and solder bead, or magnetic head); connects to the cable leading to the headset or digital counter.
- Penta contact: A gear-reduction contact in the Type AA meter that closes the circuit once every five bucket-wheel revolutions; used at high velocities to reduce the counting rate to a manageable level; not available on the Pygmy meter (Rantz et al., 1982).
- Pivot: The tungsten carbide tip at the bottom of the current meter shaft; rides in a hardened bearing cup; the single most wear-sensitive part of a mechanical current meter; must be inspected for burrs, flat spots, and magnetization before each field season (Rantz et al., 1982; U.S. Geological Survey, Office of Surface Water, 2002).
- Price Pygmy current meter: A vertical-axis cup-type current meter with a 2-inch (5.1 cm) diameter bucket wheel; designed for shallow streams (depth ≤ 1.0 ft); wading rod suspension only; rated for velocities of 0.1–4.9 ft/s (Rantz et al., 1982).
- Price Type AA current meter: The standard USGS vertical-axis cup-type current meter with a 5-inch (12.7 cm) diameter bucket wheel; rated for velocities of 0.1–25 ft/s; can be suspended from a wading rod or cable (Rantz et al., 1982).
- Rating table: A lookup table, supplied with each meter by the manufacturer and approved by the USGS, that converts bucket-wheel revolutions per second to velocity in ft/s or m/s; laminated copies must be kept in each meter’s carrying case (Rantz et al., 1982).
- Raising nut: The nut that controls the vertical position of the shaft and pivot within the yoke; must be tightened to the correct end-play setting (~0.001 in) for accurate measurements; never loosen by holding the nut and spinning the bucket wheel.
- Spin test: A field quality control check in which the bucket wheel is given a sharp spin by hand and the time for it to stop is measured; a proxy for bearing condition; minimum acceptable spin times are 40 s (Type AA) and 24 s (Pygmy) at room temperature (Rantz et al., 1982).
- Wading rod (top-setting): A graduated aluminum rod used to position the current meter at a known depth fraction (0.6 or 0.2/0.8 of total depth) below the water surface; the depth-setting mechanism is located on the handle, enabling depth changes without removing the meter from the water.
5 Safety and Hazards
- Physical: Pivot points and shaft tips are sharp; handle with care. The bucket wheel can pinch fingers if spun while adjusting the assembly.
- Chemical: Instrument oil is a mild skin irritant. Avoid eye contact.
5.1 Required PPE
6 Instrument Specifications
6.1 Price Type AA Current Meter
The Type AA is the standard USGS Price-type current meter for general streamflow measurement (Rantz et al., 1982). The bucket wheel has six conical cups (12.7 cm diameter) rotating on a vertical axis inside a brass yoke. Tungsten carbide bearings are seated in deeply recessed inverted cups that trap air and exclude water and silt. A tailpiece assembly aligns the meter with the current when suspended by cable.
| Parameter | Value |
|---|---|
| Velocity range | 0.1–25 ft/s (0.03–7.6 m/s) |
| Bucket wheel diameter | 5 in (12.7 cm) |
| Number of cups | 6 |
| Contact types | Single revolution and penta (every 5 rev) |
| Suspension | Wading rod or cable with sounding weight |
| Construction | Brass and stainless steel, chrome plated |
| Rating | USGS standard rating table (two equations joined at 1 rev/s = 2.2 ft/s) (Rantz et al., 1982) |
6.2 Price Pygmy Current Meter
The Pygmy meter is scaled to two-fifths the size of the Type AA, designed for shallow streams (Rantz et al., 1982). The contact chamber is integral to the yoke; single-revolution contact only (no penta gear). The meter has no tailpiece and cannot be used with cable suspension.
| Parameter | Value |
|---|---|
| Velocity range | 0.1–4.9 ft/s (0.03–1.5 m/s) |
| Bucket wheel diameter | 2 in (5.1 cm) |
| Number of cups | 6 |
| Contact type | Single revolution only |
| Suspension | Wading rod only (no cable suspension) |
| Construction | Brass and stainless steel, chrome plated |
| Rating | USGS standard rating table (single equation) (Rantz et al., 1982) |
6.3 Choosing Between Meters
| Condition | Meter to Use |
|---|---|
| Water depth > 1.0 ft (0.3 m) | Type AA |
| Water depth ≤ 1.0 ft (0.3 m) | Pygmy |
| Velocities expected > 4.9 ft/s | Type AA |
| Shallow stream edges (supplemental data) | Pygmy |
| Cable suspension required | Type AA |
7 Equipment and Accessories
| Item | Used With | Notes |
|---|---|---|
| Top-setting wading rod | Both | Graduated in tenths of a foot; depth-setting mechanism on handle |
| Round wading rod | Both | Simpler design; depth read from rod markings |
| Headset (earpiece) | Both | For counting clicks manually |
| AquaCalc or digital flow computer | Both | Automatically counts revolutions and computes velocity |
| Stopwatch | Both | For timed revolution counts if no digital counter |
| Sounding weight (Columbus type) | Type AA only | For cable-suspended measurements |
| Tailpiece assembly | Type AA only | Remove when using wading rod |
| Electrical cable | Both | Connects meter to counter or headset |
7.1 Maintenance Supplies
| Item | Notes |
|---|---|
| Instrument oil (light machine oil) | Supplied with meter; for pivot bearings |
| Lint-free cloth | For cleaning and drying |
| Cotton swabs (pointed) | For cleaning pivot bearing recesses |
| Small flat-head screwdriver | For contact chamber adjustments |
| Magnifying glass or loupe | For inspecting pivot points for burrs |
| Paper clip | For checking pivot magnetization |
| Bulk magnetic tape eraser | For demagnetizing pivots |
| Soft-bristle brush | For removing debris from cups |
| Compressed air (filtered) | For drying internal components |
8 Procedure
8.1 Assembly
Type AA on a wading rod:
- Remove the tailpiece assembly (used for cable suspension only).
- Slide the meter yoke onto the wading rod and secure with the thumbscrew.
- Connect the electrical cable from the binding posts on the contact chamber to the AquaCalc or headset.
- For penta contact, connect to the penta binding post; for single-revolution counting, use the single-revolution binding post.
Pygmy on a wading rod:
- Slide the pygmy yoke directly onto the wading rod and secure with the thumbscrew.
- Connect the electrical cable to the binding post. The Pygmy has only a single-revolution contact.
8.2 Pre-Field Spin Test
Perform a spin test before every field outing (Rantz et al., 1982). Conduct with the meter level, shaft vertical, in a location with no air currents.
- Hold the meter steady and use your finger or a pencil to give the bucket wheel a sharp, single spin.
- Immediately start a stopwatch.
- Stop the stopwatch when the wheel completely stops rotating.
- Record the spin time in the field notebook.
Minimum acceptable spin times (Rantz et al., 1982):
| Meter | Minimum Spin Time | Optimum Spin Time |
|---|---|---|
| Type AA | 40 s | 60–120 s |
| Pygmy | 24 s | 40–80 s |
If the spin time falls below the minimum, clean and oil the pivot before proceeding (§7.3). If the spin time remains low after cleaning and oiling, replace the pivot.
8.3 Cleaning and Oiling (Pre-Season and As Needed)
- Inspect the pivot. Remove the raising nut and carefully withdraw the shaft. Examine the pivot tip under a magnifying glass. A good pivot is highly polished, with no flat spots, pits, or scratches. Replace a damaged pivot immediately.
- Check for magnetization. Hold the pivot near a paper clip. If the clip is attracted to the pivot, the pivot is magnetized and will cause erroneously high velocity readings. Demagnetize using a bulk magnetic tape eraser.
- Oil the bearings. Apply one small drop of instrument oil to the upper shaft bearing and one drop to the pivot. Do not over-oil; excess oil attracts grit.
- Inspect the bucket wheel. Spin the wheel slowly and watch for wobble. If the upper hub wobbles more than the lower, the shaft is bent and must be replaced.
- Inspect the contact chamber. For cat-whisker contacts, check that the wire makes clean contact with the solder bead on each revolution. For magnetic-head contacts, verify the counter registers properly during the spin test.
Never loosen the raising nut by holding the nut and spinning the bucket wheel. This will bend the shaft. Always hold the bucket wheel still and unscrew the nut (U.S. Geological Survey, Office of Surface Water, 2002).
8.4 Annual Maintenance
- Disassemble the meter completely: remove bucket wheel, shaft, and pivot.
- Clean all parts with a lint-free cloth dampened with instrument oil.
- Inspect the shaft for straightness by rolling on a flat surface.
- Replace worn or damaged parts.
- Reassemble and adjust the pivot per the manufacturer’s instructions.
- Perform a spin test to verify performance.
- Record maintenance date, parts replaced, and spin test results in the instrument log (U.S. Geological Survey, Office of Surface Water, 2002).
8.5 Pivot Adjustment
- Hold the meter inverted with the pivot pointing up.
- Release the keeper screw and unscrew the pivot-adjusting nut a few turns.
- Advance the pivot-adjusting nut until it touches the yoke.
- Set end play to approximately 0.001 in – the shaft should have just barely perceptible vertical play detectable by feel.
- Tighten the keeper screw to lock the adjustment.
8.6 Storage
- Store meters in their waterproof carrying cases.
- Loosen the raising nut to relieve pressure on the pivot.
- Ensure meters are clean and lightly oiled.
- Store in a dry location away from dust and vibration.
- If a meter will not be used for more than 3 months, remove the bucket wheel assembly and store separately.
9 Rating Tables
Each meter ships with a USGS-approved standard rating table that converts bucket-wheel revolutions per second to velocity in ft/s or m/s (Rantz et al., 1982; Turnipseed & Sauer, 2010). Keep a laminated copy in each meter’s carrying case and a digital copy on the lab shared drive.
- Type AA rating table: https://prph2o.com/usgs-type-aa-current-meter/
- Pygmy rating table: https://prph2o.com/usgs-pygmy-current-meter/
9.1 Meter Calibration Verification
USGS studies have found that approximately 35% of Type AA meters and 20% of Pygmy meters in routine field use fail to meet standard rating criteria, typically because of worn pivots, bent shafts, or damaged contacts (Rantz et al., 1982). The spin test (§7.2) is the primary field check.
10 Ordering Parts and Supplies
Primary supplier for USGS-type Price AA and Pygmy current meters and replacement parts:
Performance Results Plus, Inc. Website: https://prph2o.com
10.1 Replacement Parts to Stock
| Part | Qty to Stock | Fits |
|---|---|---|
| Pivot (tungsten carbide tip) | 2 | Both |
| Shaft | 1 per type | AA or Pygmy |
| Bucket wheel assembly | 1 per type | AA or Pygmy |
| Bucket wheel hub nut | 2 | Both |
| Raising nut | 1 per type | AA or Pygmy |
| Cat-whisker wire and solder bead | 2 sets | AA |
| Instrument oil | 1 bottle | Both |
| Pivot bearing | 2 | Both |
11 Troubleshooting
| Problem | Probable Cause | Corrective Action |
|---|---|---|
| Spin time below minimum | Dirty or worn pivot; bent shaft; dry bearings | Clean pivot; inspect for burrs; oil bearings; replace pivot or shaft if damaged |
| Counter not registering | Loose cable; worn cat-whisker; dead counter battery | Check cable; replace contact wire; replace battery |
| Bucket wheel wobbles | Bent shaft or frame | Replace shaft; if hub wobbles, replace bucket wheel assembly |
| Meter reads high | Magnetized pivot; worn bearing | Demagnetize pivot; replace pivot bearing |
| Meter reads low | Excessive friction; over-oiled | Clean and re-oil; replace pivot if burred |
12 References
- Rantz et al. (1982)
- Turnipseed & Sauer (2010)
- U.S. Geological Survey, Office of Surface Water (2002)
14 Training Record
Before handling or using current meters independently, personnel must:
- Read this SOP in full
- Complete SOP-F001 Field Safety training
- Observe a trained operator assemble, spin-test, and clean both meter types
- Successfully perform assembly and spin test under supervision
- Read SOP-F301 before using meters in the field
- Be signed off by the PI or designated trainer